Chapter 13 Maintain the Instrument
13
Maintenance wizard procedures
4. Prepare for array storage: Fill an array tip cover and array electrode cover with purified water.
1
Figure 67 Array tip cover and electrode cover for the stored capillary array
Array tip cover
1
Electrode cover
2
5. Remove the capillary array:
a. Loosen the array port lock by turning it ~1/4 turn counter-clockwise until it stops. Adjust it
until the rectangular opening aligns with the rectangular head of the array tip.
Figure 68 Array port lock
Array port lock and array-port end of the capillary array
1
Retractable end of capillary array that protects the tip (slides to the right when you squeeze tab 1 and tab 2
2
toward each other)
Tab 1
3
Tab 2
4
b. Grasp the array-port end of the array and ease the tip out of the pump block.
422
1
2
3
SeqStudio
2
4
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide