Chapter 12 Run calibrations and install checks
12
Perform an install run
QV20 CRL limits—sequencing install run
Table 15 Sequencing Standards, BigDye
Passing criteria
QV20 CRL
QV20 CRL (range)
Table 16 Sequencing Standards, BigDye
Passing criteria
QV20 CRL
QV20 CRL (range)
View the spectral calibration generated by an install run
1. In any install run screen, tap Actions.
2. Tap View spectral calibration.
The spectral calibration plot is displayed.
For more information, see "View spectral calibration results"
Export sequencing install run logs or report
1. In the plot screen or the results per capillary screen, tap Actions.
2. Tap the item to export:
• Logs (ZIP)—Contains the install run report (PDF), data files (AB1), install run file (CID), and log
files.
Log file name format: InstrumentName_ChemistryType_RunDate_Timestamp.zip, for
example: Instrument1_BDTv3.1_20201021_121952.zip
• Report (PDF)—Contains the results of the install run.
Report file name format: InstrumentName_ChemistryType_RunDate_Timestamp.pdf, for
example: Instrument1_BDTv3.1_20201021_121952.pdf
3. Select the export location, then tap Next.
Note: It can take up to 15 seconds for the instrument to recognize a USB drive after the drive is
inserted in a USB port. exFAT-formatted USB drives are recommended.
346
Terminator v1.1
Instrument configuration
36-cm capillary
36-cm capillary
array POP-6
array POP-7
600
20–619
Terminator v3.1
Instrument configuration
36-cm capillary
36-cm capillary
array POP-6
array POP-7
600
40–639
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
50-cm capillary
array POP-6
600
600
20–619
20–619
50-cm capillary
array POP-6
500
600
40–539
40–639
on page
50-cm capillary
array POP-7
850
20–869
50-cm capillary
array POP-7
500
40–539
317.