Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual page 405

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

8. On the array to be installed, slide the retractor tab on the array to the right until it releases.
Figure 54 Slide the retractor tab to the right
9. Insert the array tip into the array port:
a. Slide the retractor tab to the left until the array tip reaches the array port lock.
b. Align the rectangular part of the array tip with the opening in the array port lock.
c. Insert the array tip into the array port. Press firmly to seat the array tip in the port.
Figure 55 Insert the array tip into the array port
10. Turn the array port lock ~1/4 turn clockwise to tighten it. Do not overtighten it.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
13
405

Advertisement

Table of Contents
loading

This manual is also suitable for:

A50369

Table of Contents