7. Tap OK.

Enable/disable the internal barcode reader (administrator only)

IMPORTANT! Before changing this setting, make sure no plate files are linked in the home screen.
Enable the internal barcode reader if you want to use the automated barcode workflow. See "Run plates
with the automated barcode workflow"
1. In the home screen, tap
2. Tap Enable, then tap Done.

Set the power save duration (administrator only)

This setting determines when the instrument enters power save mode. The duration you specify starts
after the latest completed run.
In power save mode, the following components are turned off: laser, oven heater, and detection cell
heater.
The instrument screens are displayed while the instrument is in power save mode. Starting a run
removes turns off power save mode.
1. In the home screen, tap
2. Tap a duration, then tap Done.