Chapter 2 Get started
2
Prepare the instrument
Check buffer fill levels
1. Open the front doors.
2. (Optional) Turn on the interior light: tap
3. Check the buffer fill levels. Verify that the buffer level is at the top of the fill line. The meniscus must
align with the fill line. Ensure that the septa on the CBC are properly seated.
1
Figure 6 Location of ABC and CBC
ABC
1
CBC
2
Figure 7 ABC fill line (the figure on the right shows the buffer level below the fill line)
ABC fill line
1
70
(Dashboard) , tap
2
1
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
, then tap On for the LED light.