In this procedure, you fill capillary array covers with purified water (if you will store the capillary array),
replace the anode buffer container (ABC) with a waste ABC, and remove the capillary array from the
instrument.
1. Prepare for array storage: Fill an array tip cover and array electrode cover with purified water.
1
Figure 40 Array tip cover and electrode cover for the stored capillary array
Array tip cover
1
Electrode cover
2
2. Open the instrument, oven, and detection block doors.
Figure 41 Location of instrument, oven, and detection cell doors
Instrument doors
1
Oven door
2
Detection cell door (with oven door open)
3
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
2
1
1
Chapter 13 Maintain the Instrument
Maintenance wizard procedures