d. Open the array header latch: pull the snap-pin connector to release it, then swing the latch
arm to the right.
Figure 44 Array header latch and array housing
Snap-pin connector on the array header latch
1
Array header
2
e. Grasp the tab on the array header and slide the array out of the instrument.
Figure 45 Tab on capillary array
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
1
2
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
13
397