Chapter 13 Maintain the Instrument
13
Maintenance wizard procedures
5. Tap Verify pouch to update the pouch information.
6. Tap Next to continue.
Change Polymer Type—Install ABC and CBC
In this procedure, you install the anode buffer container (ABC) and the cathode buffer container (CBC),
then flush the pump and channels with polymer.
1. Remove the waste ABC.
1
Figure 59 Location of ABC and CBC
Anode buffer container (ABC)
1
2. Re-install the ABC that was removed earlier, or install a new ABC. See "Install the anode buffer
container (ABC)"
3. (Optional) Install a new CBC. See "Install the cathode buffer container (CBC)"
4. Close the instrument doors.
5. Tap Verify buffer to update the buffer information.
6. Tap Flush pump and channels.
Wait ~1 minute for the flush procedure to complete.
408
on page
364.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
2
Cathode buffer container (CBC)
2
on page
367.