Chapter 13 Maintain the Instrument
13
Maintenance wizard procedures
7. Tap Verify pouch to update the pouch information.
8. Remove the ABC, discard the contents, then re-install the empty ABC in the instrument.
9. Open the oven and detection block doors.
Figure 74 Location of instrument, oven, and detection cell doors
Instrument doors
1
Oven door
2
Detection block door (with oven door open)
3
10. Loosen the array port lock by turning it ~1/4 turn counter-clockwise until it stops.