Advertisement

4. Tap Alleles to display the sizing information for alleles.
For a description of results, see "Pass/fail criteria and results for the fragment install check"
page
350.
IMPORTANT! If fewer than 15 alleles per capillary are identified, ensure that polymer has been
installed on the instrument for ≤3 days. Older polymer can cause the fragment install run to fail.
5. Tap View plot to display the plot for each capillary (Raw and Analyzed tabs are shown below).
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Chapter 12 Run calibrations and install checks
12
Perform an install run
on
349