13. Close the detection block door, then tighten the fastener.
IMPORTANT! Ensure that the fastener on the detection cell door is firmly tightened. If the fastener
is not firmly tightened, peaks may be incorrectly detected.
14. Close the oven door.
15. Tap Verify array to update the array information.
16. Tap Next to continue.
Change Polymer Type or Wash Pump and Channels—Install polymer
In this procedure, you remove the conditioning reagent that was used to wash the pump and channels.
You then install the new polymer pouch.
1. Open the instrument doors.
2. Remove the conditioning reagent pouch by pushing the polymer pouch lever down.
1
2
Figure 58 Pouch slot and lever
Slot of the lever assembly
1
Label facing the instrument
2
Lever
3
3. Wipe the pouch connector on the instrument with a moistened lab wipe.
4. Install the new polymer pouch:
a. Peel off the seal on the pouch fitting.
b. With the label facing the rear of the instrument, insert the pouch fitting into the slot on the
pump lever mechanism.
c. Push the polymer pouch lever up.
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
3
13