12
Run a spectral calibration
Note: You can also export these items from the Spectral calibration history option.
1. In the Spectral Calibration plot screen or the results per capillary screen, tap Actions.
2. Tap the item to export:
• Logs (ZIP)—Contains the spectral calibration report (PDF), data files (FSA or AB1), spectral
calibration run file (CID), and log files that can be used for troubleshooting.
Log file name format: InstrumentName_DyeSet_RunDate_Timestamp.zip, for example:
Instrument1_Z_BigDye Terminator v3.1_20201021_121952.zip
• Report (PDF)—QC report Contains the results of the spectral calibration for your records.
Report file name format: InstrumentName_DyeSet_RunDate_Timestamp.pdf, for
example: Instrument1_Z_BigDye Terminator v3.1_20201021_121952.pdf
3. Select the export location, then tap Next.
Note: It can take up to 15 seconds for the instrument to recognize a USB drive after the drive is
inserted in a USB port. exFAT-formatted USB drives are recommended.
For information on connecting to a network drive, see "Connect the software to a network drive
(software settings)"
4. When the export is complete, click Close to close the message box.

View and export the spectral calibration history

1. In the home screen, tap
2. (Optional) Tap
326
on page
443.
Actions4Maintenance4Calibration4Spectral calibration history.
Filter to filter the list by polymer, result, and/or dye set.