Chapter 5 Create a plate file in the Plate Manager software
5
Set up cloud analysis in the Plate Manager software (cloud)
Set up cloud analysis in the Plate Manager software (cloud)
You can create a plate file that specifies automatic secondary analysis in a Thermo Fisher
application. See "Secondary analysis software"
Before you set up cloud analysis, note the following:
• Set up an analysis template in the secondary analysis software if you are using VA, NGC, or MSA
software for cloud analysis. Analysis templates are optional for QC and PS software.
• In the plate file, you must enable and set up cloud analysis on the Properties tab, and specify
cloud analysis groups on the Plate tab.
• In the cloud analysis settings in the plate file, you can select an existing project or create a new
project for results in the secondary analysis software.
1. In Plate Manager software (cloud) , create a plate file. See "Set up a plate file"
2. In the Properties tab, enable the Cloud analysis option.
3. In the Cloud analysis application list, select the application.
The options displayed in the list depend on the Application type selected for the plate (above the
cloud analysis settings).
4. Select the name of a project in the secondary analysis software or create a new project. Cloud
analysis results are added to the specified project.
124
on page
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
50.
™
Connect
on page
115.