Note: A polymer pouch includes a small reserve volume that is used for the Remove Bubbles
maintenance wizard, which consumes ~350 μL of polymer. The reserve volume is sufficient to run the
wizard ~4 times (including the remove bubbles step during other maintenance wizards). If you manually
run the Remove Bubbles maintenance wizard >4 times, the volume of polymer that is available for
samples may be depleted.
For the details on messages that are displayed, see "Check consumables status"
Conditioning reagent storage, usage, and limits
Cat. No.
4393718
Conditioning Reagent Kit 3500/Flex Series
1 unit (single-use)
[1]
See the expiration date on the label. Do not use expired product.
On-instrument supported limits
Single use or expiration date
Capillary array storage and usage
WARNING! SHARP The electrodes on the load-end of the capillary array have small, blunt ends that
can lead to piercing injury.
Cat No.
A49104
Capillary array 36‑cm SeqStudio
A49106
Capillary array 50‑cm SeqStudio
A49105
Capillary array 36‑cm SeqStudio
A49107
Capillary array 50‑cm SeqStudio
On-instrument supported limits
160 injections when used with Thermo Fisher Scientific
reagents, or expiration date
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Instrument consumables handling, usage limits, and expiration
Description
Description
™
8 Flex
™
8 Flex
™
24 Flex
™
24 Flex
Chapter 1 Product information
Storage
2–8°C
After removing from storage, use the pouch within
24 hours.
Guidelines
See the expiration date on the label.
Guidelines
Capillary arrays have been verified for 160
injections when used with Thermo Fisher
Scientific reagents.
If you remove a capillary array from the instrument
for future use, see "Store a capillary array"
page
1
on page
66.
[1]
Storage
15–30°C
on
375.
27