Sae Error Messages And Actions - Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

Chapter 10 Use the instrument with the Security, Auditing, and E‑signature (SAE) v2.1 module

SAE error messages and actions

Observation
Unable to connect to SAE
server
message
The Open file from non-SAE
system Setting from SAE
server setting in the SAE
Administrator Console does
not change the software
behavior
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Possible cause
The SAE server connection
settings are incorrect.
There is a problem with
the computer on which the
SAE Administrator Console is
installed or a problem with the
network.
The computer on which the
SAE Administrator Console is
installed has a dynamic IP
address that is disconnecting
the server when the computer
is restarted.
Firewall settings are not
correctly set.
The Open file from non-
SAE system Setting from
SAE server function is not
supported in this software.

SAE error messages and actions

Recommended action
1. Check the SAE server IP address (see
"Determine the IP address for a computer
on a network"
on page
2. In the instrument Sign In screen, sign in
with a local administrator profile.
3. Set the correct IP address (in the home
screen, tap
Actions4Settings4 SAE
settings4Connection settings).
Troubleshoot computer or network problems.
Set a static IP address on the computer.
Contact Technical Support.
No action.