C
Instrument dimensions and clearance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 531
Operating specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 533
Environmental requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 533
Electrical requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 534
Wired and wireless network options and requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 535
Network and firewall requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 536
Power and communication connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 537

Instrument dimensions and clearance

Parameter
Depth
Width
Height
Weight
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Instrument footprint
~68 cm (27 inches)
~70 cm (28 inches) with doors closed
~99 cm (39 inches) with doors open
87 cm (35 inches)
~115 kg (254 lbs)
A bench that is rated for ~150 kg (331 lb) is recommended.
Instrument specifications
~25 cm (10 inches) at the rear of the
instrument to ensure adequate airflow and
cooling
~18 cm (7 inches) from wall
~36 cm (14 inches) from other instruments
or computers
~60 cm (24 inches) from the top of the
instrument
~150 cm (60 inches) from the bench top
Recommended clearance