3. To export the list of calibrations that have been run on the instrument, tap Actions4Export
report.
The spectral calibration history report contains the results of all spectral calibrations performed on
the instrument.
Report file name format:
InstrumentName_SpectralHistoryReport_ReportGenerationDate_Timestamp.pdf, for
example: Instrument1_SpectralHistoryReport_20201124_162247.pdf
4. To view the details of an individual spectral calibration, tap a dye set, then tap View.
5. To export the log or report for the individual spectral calibration, tap Actions, then tap the item to
export (these options generate the same files you can generate at the end of a spectral calibration):
• Logs (ZIP)—Contains the spectral calibration report (PDF), data files (FSA or AB1), spectral
calibration run file (CID), and log files that can be used for troubleshooting.
Log file name format: InstrumentName_DyeSet_Timestamp.zip, for example:
Instrument1_Z_BigDye Terminator v3.1_20201021_121952.zip
• Report (PDF)—
Contains the results of the spectral calibration for your records.
Report file name format: InstrumentName_DyeSet_date_Timestamp.pdf, for example:
Instrument1_Z_BigDye Terminator v3.1_20201021_121952.pdf
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
Chapter 12 Run calibrations and install checks
Run a spectral calibration
12
327