Note: You may notice a tiny droplet of polymer inside the fitting (residual from the pouch filling
process). This is not expected to cause any performance issues.
6. With the RFID label facing toward the instrument, slide the pouch fitting onto the slot of the lever
assembly. Push the lever up to snap the pouch into the connector end of the instrument pump.
1
2
Slot of the lever assembly
1
Label facing the instrument
2
Lever
3
Note: The RFID label must face the instrument (away from you) to ensure that the RFID
information is read accurately by the instrument.
7. Follow the instructions in the wizard to finish the procedure.
Store partially used polymer
If you remove a polymer pouch for storage (2–8°C), place a pouch cap (Cat. No. 4462785) onto the
pouch. Install an empty pouch or conditioning reagent pouch on the instrument to prevent desiccation
of any residual polymer on the connector. Follow the instructions in the maintenance wizards to ensure
proper installation of the polymer pouch.
Change and store a capillary array (wizards)
WARNING! SHARP The electrodes on the load-end of the capillary array have small, blunt ends that
can lead to piercing injury.
IMPORTANT! Wear appropriate protection, including gloves, laboratory goggles, and coat whenever
you work with the fluids used on this instrument, or parts that may come into contact with these fluids.