Chapter 3

Installation

Unpacking Inspection

Orion 8030cX Silica Analyzer has been carefully tested
and packaged before leaving the factory.
Open the inner and outer packing and check the product
and accessories according to the Packing Checklist. If any
components are missing or damaged, please contact
Thermo Fisher or its representative office immediately.
Open the outer package and take out the foam supports at
the corners of the inner package.
Open the inner packing and check the product and
accessories according to the Packing checklist.
1
Outer Box
2
Inner Box
3
Accessories Box
4
Upper Foam Insert
8
| Chapter 3 Installation
5
Corner Foam Blocks
6
Bottom Foam Insert
Box Reinforcement
7
Sleeves
Requirements for Instrument
Installation
Basic Requirements for Instrument Installation
Instructions for instrument handling:
CAUTION: When equipment exceeding 40
lbs. should be lifted with the help of two
persons or follow any heavy lifting policies
your employer in place.
It is recommended that two people work together
to lift and handle the instrument.
In the process of lifting the instrument, the
instrument door should not be used as the lifting
point, otherwise the door and sealing elements
could become damaged.
Selection of Installation Location
Please refer to the specifications for the temperature and
relative
humidity
requirements
environment.
It is recommended to install the instrument in a shed,
protective case, or analysis room to avoid extreme
weather and freezing reagents. For the best results, it is
recommended
to
install
temperatures of 10-30 ℃ and maximum humidity < 95%
(no condensation) room.
Orion 8030cX Silica Analyzer has reached IP65 protection
level and can be installed outdoors. However, the
installation location of the instrument cannot be flooded,
frozen, or overheating due to direct sunlight. If the ambient
temperature drops below freezing point, the reagent
solutions may freeze and stop analysis.
8030cX Silica Analyzer
of
the
installation
the
silica
analyzer
at