Pre-Run Check Messages And Plate Alerts During A Run; Pre-Run Checks When A Plate Run Starts Or A Re-Injection Or A Cancelled Injection Is Added To The Run Queue; Pre-Run Checks Before Each Injection Or When A Paused Run Is Resumed - Thermo Scientific appliedbiosystems SeqStudio Flex Series User Manual

Genetic analyzer with instrument software v1.0
Table of Contents

Advertisement

Chapter 6 Start and monitor a run
6

Pre-run check messages and plate alerts during a run

Pre-run check messages and plate alerts during a run

Pre-run checks are performed at the following times during a run:
• When a plate run starts
• When a re-injection is added to the Run queue
• Before each injection
• When a paused run is resumed
• When a cancelled injection is added back to the Run queue
A partial list of pre-run checks is shown below.
• Checks for valid spatial and spectral calibrations.
• Checks for a passing install run.
• Checks consumables expiration, volume, and number of remaining injections.
• Determines if adequate storage space is available for results.
• Determines if the plate has been installed in the drawer for more than 24 hours.
• Checks that the save locations specified in the plate file are available and writable.
• If you run using the automated barcode workflow, performs additional checks. See "Run plates
with the automated barcode workflow"
Pre-run checks when a plate run starts or a re-injection or a cancelled
injection is added to the Run queue
Pre-run checks are performed when a plate run starts or when a re-injection or a cancelled injection
is added to the Run queue. If these pre-run checks do not pass, a message screen is displayed. Two
message types are possible:
• Pre-run check failure message—The error condition must be corrected before the run can start.
• Pre-run check warning message—For information only, tap OK to start the run.
Examples of pre-run check failure messages include
array",
"Plate type of the installed plate does not match the plate
Examples of pre-run check warning messages include
exceeded 24 hours on

Pre-run checks before each injection or when a paused run is resumed

Pre-run checks are also performed before each injection or when a paused run is resumed. If these
pre-run checks do not pass, a
To display the reason for the alert, tap the
displayed. For more information, see the following sections:
• "View alert and notification details"
• "Plate alert troubleshooting"
146
on page
instrument",
"Network drive save location is not
(Plate alert) symbol is displayed at the top of a plate position.
!
(Plate alert) symbol. The Notifications screen is
!
on page 149
on page 507
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
147.
"No spatial calibration available for capillary
file".
"Injections on plate will exceed or have
writable".