Chapter 13 Maintain the Instrument
13
Maintenance wizard procedures
7. Tap Wash pump and channels.
IMPORTANT! After you tap Wash pump and channels you cannot cancel the procedure. The
wash procedure takes ~45 minutes to complete.
Shut Down Instrument—Clean the instrument interior
In this procedure, you empty the waste ABC and fill it with purified water, clean the instrument interior,
then manually flush the water trap. Refer to the procedures below for more information.
1. Open the instrument doors.
2. Remove the ABC and discard the waste.
3. Fill the empty ABC with water, then re-install it.
Note: Check the fluid level in the ABC every 2 weeks. Refill as needed.
4. Leave the installed conditioning reagent pouch installed in the instrument during storage.
5. Flush the water trap. See "Flush the water trap (pump trap)"
6. Clean the instrument interior. See "Clean the instrument"
7. Tap Next to continue.
Shut Down Instrument—Power off instrument
In this procedure, you power off the instrument.
1. Close the instrument doors.
2. Press the On/Off button on the front panel to power off the instrument.
1
Figure 73 Front panel of instrument
On/Off button
1
426
SeqStudio
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
on page
379.
on page
361.