Lcd Module Replacement; Figure 7-13. Replacing The Lcd Module - Thermo Scientific 43iQTL Instruction Manual

Trace level sulfur dioxide analyzer
Table of Contents

Advertisement

LCD Module
Use the following procedure to replace the LCD module.
Replacement
Equipment required:
Wrench, 1/4
1. Turn instrument OFF and unplug the power cord.
2. Gripping from the top corners of the front panel and pull outwards.
3. Unfasten four nuts
(Figure
7–13).
Figure 7–13. Replacing the LCD Module
4. Remove cover.
5. Unplug LCD cables from backside of board.
6. Pull board off the standoffs.
43iQTL Instruction Manual
Thermo Scientific