3.1.2
Model 20-35-NM-F
SW1 and SW2 Settings are used to delay the indication of alarm conditions. This will prevent
alarms due to material momentarily moving the probe.
Table 3-2: SW1 and SW2 Settings
SW1
SW2
B
B
A
B
B
A
A
A
SW3 determines whether the alarm functions of the control unit respond like a Flow or Level
application.

Table 3-3: SW3 Settings

SW3
Mode
A
Flow
B
Level
SW5 determines whether the alarm functions of the control unit respond like a High or Low Level
application.

Table 3-4: SW5 Settings

SW5
Mode
A
High Level
B
Low Level
SW4 defines the instrument electrical power supply
Table 3-5: SW4 Selector Setting
SW4
Power Supply
Position
115VAC
Position
230VAC
3-4
Switches Setting
Delay in Seconds
1
2
4
6

Advertisement