SCPI Command Reference
E4980A/AL commands
Equivalent key
No equivalent key is available on the front panel.
:SYSTem:PERSona:MANufacturer:DEFault
Syntax
:SYSTem:PERSona:MANufacturer:DEFault
Description
This command sets the instrument's manufacturer to "Keysight Technologies"
which is the default manufacturer setting.
Equivalent key
No equivalent key is available on the front panel.
:SYSTem:PRESet
Syntax
:SYSTem:PRESet
Description
Resets the instrument settings and correction data. (No Query)
Equivalent key
[Preset] - CLEAR SET&CORR - OK
:SYSTem:RESTart
Syntax
:SYSTem:RESTart
Description
Reboots the instrument immediately. (No Query)
This is available only for serial prefix number below:
MY464/SG464 and below for E4980A
MY542/SG542 and below for E4980AL
Equivalent key
No equivalent key is available on the front panel.
:SYSTem:TIME
Syntax
:SYSTem:TIME <hour>,<minute>,<second>
:SYSTem:TIME?
Description
Sets the date in the internal clock.
Parameter
Range
Unit
Resolution
372
<hour>
0 to 23
hours
1
Keysight E4980A/AL Precision LCR Meter