Tests and Adjustments
Adjustments
Figure 3-19
Equipment Setup 3 for the Receiver Adjustment
EE Default Adjustment
This sets the EEPROM data to their default values.
Procedure
1. Press UTILITY
2. On the dialog box, select Initialize rather than Adjust/Verify because an
3. Follow the instructions and prompts as they are displayed.
IF Response Adjustment (Available Only with Option S93093/4A and
or S93900/1/2/4/5/7A Installed)
In A models these options were 090 and 093 or 094.
The IF Response adjustment is required when upgrading to Option 090, 093
and or 094. In addition, this adjustment must be performed each time a new
configuration is set up or if cables are changed:
Procedure
1. Press UUTILITY
2. Follow the instructions and prompts as they are displayed.
3-50
System
click EE Default Adjustment.
adjustment is typically unnecessary.
System
click IF Response Adjustment.
Service
Adjustment Routines
, then
, then
Service
, then
, then
Keysight N5224B/25B Service Guide
, then
Adjustment Routines
, then