Default And Advanced Measurement Parameters; Saving And Loading Setups; Saving Setups - Keysight N4372E User Manual

Lightwave component analyzer
Table of Contents

Advertisement

5
User Interface Reference

Default and Advanced Measurement Parameters

Saving and Loading Setups

Saving setups

80
Use the default parameters for EO measurements of devices with a simple
stimulus and response path. This includes devices like sources, receivers
or amplifiers.
Use the advanced parameters for more comprehensive control of your
measurements.
A setup contains the current state of the LCA user interface and optionally
also the current state of the network analyzer. Setups are stored as
XML-files or as compressed XML-files.
The settings on the "Tools" tab are not saved in the setup, since the
settings shown on this tab always reflect the current state of the LCA
hardware. After loading the setup the LCA system is configured to press
"Start" from the LCA user interface.
There are two ways of saving setups.
First saving method
Click
to save the current setup.
If user calibration mode is set to "Measure" (only possible for OE and
OO measurements), you are asked to specify the electrical calibration
set to be saved with the setup.
This is the electrical calibration set to be used during user calibration
measurements.
Only a link to the electrical calibration set is saved, not the calibration
data itself.
For more details, see
page 54 or
Measuring Optical-Optical (OO) Devices
If User Calibration is set to "Load" (only possible for OE and OO
measurements), you will be asked to specify which user calibration file
should be loaded.
The filename of the calibration data is saved with the setup, not the
calibration data itself.
For more details, see
page 54 or
Measuring Optical-Optical (OO) Devices
Keysight N4372E Lightwave Component Analyzer User's Guide
Measuring Opto-Electrical (OE) devices
Measuring Opto-Electrical (OE) devices
on
on page 70.
on
on page 70.