User Interface Reference

Default and Advanced Measurement Parameters

Saving and Loading Setups

Saving setups

64
Use the default parameters for EO, OE and OO
measurements of devices with a simple stimulus and
response path. This includes devices like sources, receivers
or amplifiers.
Use the advanced parameters for more comprehensive
control of your measurements. For example, to use the
optical output as the DC light source to an optical modulator
as part of an EO measurement.
A setup contains the current state of the LCA user interface
and optionally also the current state of the network analyzer.
Setups are stored as XML- files or as compressed XML- files.
The settings on the "Tools" tab are not saved in the setup,
since the settings shown on this tab always reflect the
current state of the LCA hardware. After loading the setup
the LCA system is configured to press "Start" from the LCA
user interface.
There are two ways of saving setups.
First saving method
Click
to save the current setup.
• If user calibration mode is set to "Measure" (only possible
for OE and OO measurements), you are asked to specify
the electrical calibration set to be saved with the setup.
This is the electrical calibration set to be used during
user calibration measurements.
Only a link to the electrical calibration set is saved, not
the calibration data itself.
For more details, see
Devices" on page 86
Devices" on page
• If User Calibration is set to "Load" (only possible for OE
and OO measurements), you will be asked to specify
which user calibration file should be loaded.
Keysight N4375D Lightwave Component Analyzer, User's Guide
"Measuring Opto- Electrical (OE)
or
"Measuring Optical- Optical (OO)
99.