Summary of Contents for Agilent Technologies 4263B

  • Page 1 Keysight Agilent HP 4263B Manual Get Pricing & Availability at ApexWaves.com Call Today: 1-800-915-6216 Email: sales@apexwaves.com https://www.apexwaves.com/meters/keysight-technologies/lcr-meters/4263B...
  • Page 2 Agilent 4263B LCR Meter Operation Manual SERIAL NUMBERS Agilent Part No. 04263-90050 Printed in JAPAN March 2003 Sixth Edition...
  • Page 3 Notice...
  • Page 4: Manual Printing History

    4263B Manual Printing History : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :...
  • Page 5 4263B Safety Summary WARNINGS The Agilent Technologies assumes no liability for the customer's failure to comply with these requirements. Note Ground The Instrument DO NOT Operate In An Explosive Atmosphere Keep Away From Live Circuits DO NOT Service Or Adjust Alone...
  • Page 6: Certi Cation

    4263B Certi cation Warranty General Information Limitation Of Warranty No other warranty is expressed or implied. Agilent Technologies speci cally disclaims the implied warranties of merchantability and tness for a particular purpose.
  • Page 7: Exclusive Remedies

    4263B Exclusive Remedies The remedies provided herein are buyer's sole and exclusive remedies. Agilent Technologies shall not be liable for any direct, indirect, special, incidental, or consequential damages, whether based on contract, tort, or any other legal theory. Assistance...
  • Page 8: Safety Symbols

    4263B Safety Symbols Warning Caution Note...
  • Page 9 4263B Herstellerbescheinigung < Manufacturer's Declaration < viii...
  • Page 10: Contents Of This Manual

    4263B Contents of this Manual Chapter 1 Chapter 2 Chapter 3 Chapter 4 Chapter 5 Chapter 6 Chapter 7 Chapter 8 Chapter 9...
  • Page 11 4263B Appendix A Appendix B Appendix C...
  • Page 12 Contents 1. Getting Started Contents-1...
  • Page 13 2. Operating the 4263B 3. Function Reference and Technical Information Contents-2...
  • Page 14 Contents-3...
  • Page 15 4. Remote Operation (To Control from a Computer) Contents-4...
  • Page 16 5. GPIB Reference j j j f j g f j g j j g f j g f j g f j g numeric value < > f j g j j g f j g j j g f j g numeric value <...
  • Page 17 j j g numeric value < > j j g numeric value numeric value < > < > f j j g f j j g j j g numeric value < > j j g j j g numeric value <...
  • Page 18 numeric value < > numeric value < > 6. Application Measurement 7. Impedance Measurement Basics Contents-7...
  • Page 19 8. Speci cations 9. Maintenance Contents-8...
  • Page 20 A. Manual Changes Contents-9...
  • Page 21 B. Handler Interface Installation C. Overload/No-Contact Operations Messages Index Contents-10...
  • Page 22 Figures Contents-11...
  • Page 23 Contents-12...
  • Page 24 Tables Contents-13...
  • Page 25 Contents-14...
  • Page 26: Getting Started

    Getting Started Getting Started 1-1...
  • Page 27 4263B Overview Features 1-2 Getting Started...
  • Page 28 4263B Accessories Available Note Options Available Getting Started 1-3...
  • Page 29 4263B Guided Tour of Panels Front Panel Display LINE Switch Chassis Terminal UNKNOWN Terminals Caution DC Bias Key DC Bias ON/OFF Indicator Measurement Time Key Average Key Show Setting Key - (Shift) Current / Voltage Monitor (I&V Mon) Key -...
  • Page 30 4263B Measurement Parameter Key (Shift) Deviation Measurement Mode (1Mode) Key - Frequency Key Display Mode Key Level Key Bias Setup Key Trigger Mode Key Delay Key Trigger Key Local Key Address Key Recall Key Save Key Primary Parameter Upper Limit (Pri High) Key -...
  • Page 31 4263B 0 Key Key Lock Key . Key Reset Key Con guration Key 3 Key / (Shift) Cable Key 2 Key Contact Check Key 1 Key Comparator Key 4 Key Open Key 5 Key Short key 6 Key Load Key 7 Key...
  • Page 32 4263B Display Character Display Area Annunciator Measurement Time Trigger Hold Range Load On Comparator On Contact Check Talk Only Remote Key Lock Measurement Settings Shift Getting Started 1-7...
  • Page 33 4263B Rear Panel Caution GPIB Interface gpib External DC Bias Terminal LINE Fuse Holder Handler Interface External Trigger Terminal Serial Number Plate LINE Voltage Selector Power Cord Receptacle 1-8 Getting Started...
  • Page 34: Incoming Inspection

    4263B Incoming Inspection Warning To avoid hazardous electrical shock, do not turn on the 4263B when there are signs of shipping damage to any portion of the outer enclosure (for example, covers, panel, or display) Providing clearance to dissipate heat at installation site...
  • Page 35: Power Cable

    4263B Power Cable Warning For protection from electrical shock, the power cable ground must not be defeated. The power plug must be plugged into an outlet that provides a protective earth ground connection. 1-10 Getting Started...
  • Page 36 4263B Figure 1-1. Power Cable Supplied Getting Started 1-11...
  • Page 37: Preparation For Turning On

    4263B Preparation for Turning On Setting the Line Voltage Power Line Voltage : Power Line Frequency : Power Consumption : Fuse Figure 1-2. Voltage Selector and Fuse Table 1-1. Line Voltage Selection Voltage Selector Power Line Required Fuse Voltage 1-12 Getting Started...
  • Page 38 4263B Turning On the 4263B and Setting the Line Frequency Note Getting Started 1-13...
  • Page 39: Using The Front-Panel Keys

    4263B Using the Front-Panel Keys Note Direct Execution Type Keys Toggle Type Keys Selection Type Keys 1-14 Getting Started...
  • Page 40 4263B Value Setup Type Keys Numeric Keys Getting Started 1-15...
  • Page 41 4263B Note Minimum and Maximum Keys Up/Right, Down/Left Arrow Keys 1-16 Getting Started...
  • Page 42 4263B Back Space Key Getting Started 1-17...
  • Page 43: Basic Operation

    4263B Basic Operation Caution Note 1-18 Getting Started...
  • Page 44 4263B Connecting a Test Fixture Figure 1-3. Connecting a Test Fixture Getting Started 1-19...
  • Page 45 4263B Resetting the 4263B 1-20 Getting Started...
  • Page 46 4263B Selecting the Measurement Parameter Primary Secondary Note Getting Started 1-21...
  • Page 47 4263B Setting the Test Signal Frequency Selecting the Measurement Range Hold Range Auto Range mode Hold Range Hold Range mode 1-22 Getting Started...
  • Page 48 4263B Setting the Test Signal Level Matching the Cable Length Getting Started 1-23...
  • Page 49 4263B Performing the OPEN Correction 1-24 Getting Started...
  • Page 50 4263B Performing the SHORT Correction Getting Started 1-25...
  • Page 51 4263B Performing the LOAD Correction 1-26 Getting Started...
  • Page 52 4263B Getting Started 1-27...
  • Page 53 4263B 1-28 Getting Started...
  • Page 54 Operating the 4263B Measurement Con guration Selecting Measurement Time Mode Setting the Averaging Rate Operating the 4263B 2-1...
  • Page 55 4263B Setting Trigger Delay Time Setting Contact Check Cont Chk Note 2-2 Operating the 4263B...
  • Page 56 4263B Setting the Beeper Mode Setting the Level Monitor Mode Operating the 4263B 2-3...
  • Page 57: Triggering A Measurement

    4263B Triggering a Measurement Trigger Note Trigger Applying the DC Bias DC Bias ON/OFF Using the Internal DC Bias Source 2-4 Operating the 4263B...
  • Page 58: Using The Comparator Function

    4263B Using an External DC Bias Source Using the Comparator Function Comprtr On Operating the 4263B 2-5...
  • Page 59 4263B Note Display Con guration Displaying Deviation Data Setting the Reference Value 2-6 Operating the 4263B...
  • Page 60 4263B Selecting the Deviation Display Mode Operating the 4263B 2-7...
  • Page 61 4263B Changing the Measurement Settings Display Mode 2-8 Operating the 4263B...
  • Page 62 4263B Setting the Display Digit and Display Mode Locking Out the Front Panel Keys Operating the 4263B 2-9...
  • Page 63: Setting The Gpib Address

    4263B Returning the Local Mode (Exiting the Remote Mode) Setting the GPIB Address Saving and Recalling Instrument Settings 2-10 Operating the 4263B...
  • Page 64: Printing Measurement Data

    4263B Note Printing Measurement Data Operating the 4263B 2-11...
  • Page 65 4263B Testing the 4263B Performing a Self-Test Testing the Front Panel Keys' Functionality Note 2-12 Operating the 4263B...
  • Page 66 4263B Operating the 4263B 2-13...
  • Page 67: If You Have A Problem

    4263B If You Have a Problem If the Display is Blank and the 4263B Appears Dead If an Error Message is Displayed If the 4263B does not Accept Any Key Input Key Lock 2-14 Operating the 4263B...
  • Page 68 Function Reference and Technical Information Function Reference and Technical Information 3-1...
  • Page 69: Front Panel

    4263B Front Panel Figure 3-1. Front Panel Note Display Meas Time Trigger Hold Range Load On Comprtr On Cont Chk Talk Only Key Lock Shift 3-2 Function Reference and Technical Information...
  • Page 70: Line Switch

    4263B LINE Switch Note Note Chassis Terminal UNKNOWN Terminals Caution Function Reference and Technical Information 3-3...
  • Page 71 4263B DC Bias Key Note Measurement Parameter Key Primary Secondary 3-4 Function Reference and Technical Information...
  • Page 72 4263B Note Deviation (1) Mode Key 1ABS mode M easuredV alue ReferenceV alue 1% mode M easuredV alue ReferenceV alue ReferenceV alue Note Measurement Time Key Average Key Function Reference and Technical Information 3-5...
  • Page 73 4263B Frequency Key Display Mode Key Data Comprtr Digit Measurement Settings Display Key Level Monitor Key 3-6 Function Reference and Technical Information...
  • Page 74 4263B Level Key Note DC Bias Setup Key Auto/Hold Range Key Hold Range Function Reference and Technical Information 3-7...
  • Page 75 4263B Range Setup Key Table 3-1. Measurement Range Selection Range Setup Optimum Measurement Range Measurable Range  j j j j  <  j j j j  <  j j j j  <  j j j j  <...
  • Page 76: Local Key

    4263B Trigger Mode Key Trigger Delay Key Note Local Key Function Reference and Technical Information 3-9...
  • Page 77 4263B Address Key stat data1 data2 comp1 comp2 < > < > < > < > < > stat < > data1 < > data2 < > comp1 < > comp2 < > Save Key Caution Recall Key Note 3-10 Function Reference and Technical Information...
  • Page 78 4263B Comparator Limit Keys Left/Down Arrow Key and Up/Right Arrow Key, 0,1,..,9, (Point), (Minus) Keys, . . . Enter Key Shift Key Shift Engineering Units Key Function Reference and Technical Information 3-11...
  • Page 79: Back Space Key

    4263B Back Space key Minimum Key Maximum Key 3-12 Function Reference and Technical Information...
  • Page 80 4263B Open Key OpenMeas MeasVal Exit j j g Function Reference and Technical Information 3-13...
  • Page 81 4263B Short Key ShortMeas MeasVal Exit j j g Note 3-14 Function Reference and Technical Information...
  • Page 82 4263B Load Key On/O Load On Load On LoadMeas Corval PrmSlct RefEnt MeasVal Exit Exit Function Reference and Technical Information 3-15...
  • Page 83 4263B Comparator Key Comptr On Destination Condition Note 3-16 Function Reference and Technical Information...
  • Page 84 4263B Contact Check Key Measurement Contact Check Range Setup Available Range < < < < < < < < < <  j j <  j j < < < Note Function Reference and Technical Information 3-17...
  • Page 85 4263B Cable Key Cable Length Reference Plane Available Frequency Range Note Key Lock Key Note 3-18 Function Reference and Technical Information...
  • Page 86 4263B Reset Key Item Reset key Stored in Function Reference and Technical Information 3-19...
  • Page 87 4263B Con guration Key Note Item Result Error Code 3-20 Function Reference and Technical Information...
  • Page 88: Rear Panel

    4263B Rear Panel Figure 3-2. Rear Panel External Trigger Terminal Figure 3-3. Required External Trigger Pulse Speci cation Function Reference and Technical Information 3-21...
  • Page 89 4263B External DC Bias Terminal Caution LINE Fuse Holder LINE Voltage Selector Table 3-3. Line Voltage selection Voltage Selector Line Voltage Required Fuse Serial Number Plate Power Cord Receptacle Power Code 3-22 Function Reference and Technical Information...
  • Page 90: Handler Interface

    4263B Handler Interface Speci cation Note Figure 3-4. Pin Assignment of Handler Interface Connector Function Reference and Technical Information 3-23...
  • Page 91 4263B Table 3-4. Pin Assignment of Handler Interface Connector Pin No. Signal Name Description E11 E12 E13 E14 E15 E20 E 313 3-24 Function Reference and Technical Information...
  • Page 92 4263B Figure 3-5. Timing Diagram Note Function Reference and Technical Information 3-25...
  • Page 93: Gpib Interface

    4263B GPIB Interface Table 3-5. GPIB Interface Capability Cord Function 3-26 Function Reference and Technical Information...
  • Page 94: Technical Information

    4263B Technical Information Overall Impedance Measurement Theory Figure 3-6. Simpli ed Model of Impedance Measurement Figure 3-7. Vector Representation of Impedance Function Reference and Technical Information 3-27...
  • Page 95 4263B Note Figure 3-8. Relationship Between Impedance and Admittance Figure 3-9. Vector Representation of Admittance 3-28 Function Reference and Technical Information...
  • Page 96 4263B Note Figure 3-10. Relationship between Measurement Parameters Function Reference and Technical Information 3-29...
  • Page 97 4263B Transformer Parameters Measurement Note Figure 3-11. Basic Transformer Measurement Setup 3-30 Function Reference and Technical Information...
  • Page 98 4263B Warning While H terminal of the 4263B is connected to a terminal of the transformer, high-voltage can be induced at the other terminals. For safe operation, when connecting or disconnecting a transformer, connect the transformer leads in the following order:...
  • Page 99 4263B Test Current Level Measurement Range Setting Figure 3-12. Test Signal for DCR Measurement 3-32 Function Reference and Technical Information...
  • Page 100 4263B DC Resistance Value < R DUT Test Current Transient Trigger Delay Delay Time in DC Resistance Setting (Td Measurement (Td < Function Reference and Technical Information 3-33...
  • Page 101 4263B Figure 3-13. Test Signal Transient in DC Resistance Measurement 3-34 Function Reference and Technical Information...
  • Page 102 Remote Operation (To Control from a Computer) Note BASIC Programming Techniques BASIC Language Reference Beginner's Guide to SCPI BASIC Interfacing Techniques Tutorial Description of the General Purpose Interface Bus Condensed Description of the General Purpose Interface Bus Remote Operation (To Control from a Computer) 4-1...
  • Page 103: Input/Output Statements

    4263B Getting Started Input/Output Statements Reading the GPIB Address Sending a Remote Command 4-2 Remote Operation (To Control from a Computer)
  • Page 104: Returning To Local Mode

    4263B Returning to Local Mode Query Commands Getting Data from the 4263B Remote Operation (To Control from a Computer) 4-3...
  • Page 105: Remote Operation

    4263B Remote Operation Step 1; Resetting Step 1; Initiating trigger system Step 1; Impedance measurement Step 1; Primary Parameter: CS Step 1; Secondary Parameter: D Step 1; Test Signal Frequency: 100 Hz Step 2; Trigger Source: Bus Step 2; Triggering Step 3;...
  • Page 106 4263B To Set Up the 4263B To Reset the 4263B Note To Set the Power Line Frequency To Match Cable Length of the Test Fixture To Select the Measurement Parameter Note Remote Operation (To Control from a Computer) 4-5...
  • Page 107 4263B 4-6 Remote Operation (To Control from a Computer)
  • Page 108 4263B To Select the Test Signal Frequency To Select the Test Signal Level To Select Measurement Range Remote Operation (To Control from a Computer) 4-7...
  • Page 109 4263B To Apply a DC Bias To Perform Correction Note 4-8 Remote Operation (To Control from a Computer)
  • Page 110 4263B To Select Measurement Time Mode Remote Operation (To Control from a Computer) 4-9...
  • Page 111 4263B To Set the Averaging Rate To Set Trigger Delay Time To Set Beeper Mode To Lock Out the Front Panel Keys To Check Contact Integrity at the Test Fixture To Use the Comparator Function 4-10 Remote Operation (To Control from a Computer)
  • Page 112 4263B f j g To Display a Deviation Measurement f j g To Wait Until Previously Sent Commands are Completed Remote Operation (To Control from a Computer) 4-11...
  • Page 113 4263B To Get the Current Instrument Settings To Save and Recall Instrument Settings 4-12 Remote Operation (To Control from a Computer)
  • Page 114 4263B To Trigger a Measurement Follow the procedure below to perform successive measurements automatically (initial setup). :TRIG:SOUR :INIT:CONT Two methods to perform a measurement at your desired time: Triggering the instrument at your desired time :TRIG:SOUR :INIT:CONT Command Query response...
  • Page 115: Waiting For Completion Of Measurement (Detecting Completion Of Measurement)

    4263B Waiting For Completion Of Measurement (detecting completion of measurement) *SRE :STAT:OPER:ENAB Figure 4-2. SRQ generation sequence (when measurement nishes) 4-14 Remote Operation (To Control from a Computer)
  • Page 116: Sample Program

    4263B Sample program Figure 4-3. Detecting the completion of measurement using SRQ Remote Operation (To Control from a Computer) 4-15...
  • Page 117: Reading Out Measured Result

    4263B Reading Out Measured Result Available trigger mode Readout procedure *TRG *TRG :FETC? :FETC? Note *TRG 4-16 Remote Operation (To Control from a Computer)
  • Page 118: Reading Out Measured Result Using *Trg Command

    4263B Reading out measured result using *TRG command *TRG :TRIG:SOUR *TRG *TRG Remote Operation (To Control from a Computer) 4-17...
  • Page 119 4263B Figure 4-4. Reading out the measured result in ASCII transfer format by using the *TRG command 4-18 Remote Operation (To Control from a Computer)
  • Page 120 4263B Note Remote Operation (To Control from a Computer) 4-19...
  • Page 121 4263B Figure 4-5. Reading out the measured result in binary transfer format using *TRG command 4-20 Remote Operation (To Control from a Computer)
  • Page 122: Reading Out Measured Result Using :Fetc? Command

    4263B Reading out measured result using :FETC? command :FETC? Note :TRIG :FETC? :FETC? :FETC? Remote Operation (To Control from a Computer) 4-21...
  • Page 123 4263B Figure 4-6. Reading out the measured result in ASCII transfer format using the :FETC? command 4-22 Remote Operation (To Control from a Computer)
  • Page 124 4263B :FETC? Note Remote Operation (To Control from a Computer) 4-23...
  • Page 125 4263B Figure 4-7. Reading out measured result in binary transfer format using :FETC? command 4-24 Remote Operation (To Control from a Computer)
  • Page 126 4263B To Retrieve Data Eciently To Use a Data Bu er Remote Operation (To Control from a Computer) 4-25...
  • Page 127: Other Features

    4263B Other Features To Test the 4263B To Report the Instrument's Status Enable Measuring bit of Operation Status Register Enable Operation Status Register Summary bit Wait until the Operation Status Register Summary bit is set Clears status byte register Sets Command Error Bit and Execution Error Bit...
  • Page 128 4263B Clear the SRQ BIt Asks contents of the Standard Event Status Register Asks to output error number and message Remote Operation (To Control from a Computer) 4-27...
  • Page 129 4263B If You Have a Problem If the 4263B Hangs Up When You Send the :ABORt Command 4-28 Remote Operation (To Control from a Computer)
  • Page 130: Gpib Commands

    GPIB Reference gpib GPIB Commands Common Commands Common Command List < > < > < > < > GPIB Reference 5-1...
  • Page 131: Subsystem Commands

    4263B Subsystem Commands command tree Subsystem Command List COMMAND PARAMETER NOTE ABORt CALCulate{1|2} :FORMat {REAL|MLINear|CP|CS|LP|LP| IMAGinary|PHASe|D|Q|REAL\LP} :LIMit :BEEPer :CONDition {FAIL|PASS} < > [:STATe] :CLEar :FAIL? :LOWer < > [:DATA] < > :STATe < > :STATe :UPPer < > [:DATA] <...
  • Page 132 4263B Subsystem Command List (continued) COMMAND PARAMETER NOTE [SENSe] :AVERage < > :COUNt < > [:STATe] :CORRection :CKIT < > < > :STNdard3 :COLLect [:ACQuire] STANdard{1|2|3} :METHod {REFL2|REFL3} :DATA? STANdard{1|2|3} < > [:STATe] :FIMPedance < > :APERture [MS|S] :CONTact <...
  • Page 133 4263B Concept of Subsystem Command Tree root command root path Message Terminator < > Colon (:) Semicolon (;) Figure 5-1. Proper Use of the Colon and Semicolon 5-4 GPIB Reference...
  • Page 134: Program Message Syntax

    4263B Program Message Syntax Case Program Message Terminator program message terminators < > < > < >< > < > Subsystem Command Syntax Common Command Syntax Parameters < > parameter parameter parameter GPIB Reference 5-5...
  • Page 135: Parameter Types

    4263B Parameter Types numeric value < > numeric value < > numeric value < > < > < > sensor function data handle < > < > sensor function < > using single quote using double quote Units 5-6 GPIB Reference...
  • Page 136: Multiple Messages

    4263B Table 5-1. Usable Units De nition Mnemonic Usable Unit Multiple Messages Query and Response Message Syntax < >< > GPIB Reference 5-7...
  • Page 137 4263B Command Reference Notations :CALClate :LIMit:LOWer[:DATA] < > :CALClate :LIMit:LOWer:STATe < > f j g < > 5-8 GPIB Reference...
  • Page 138: Abort Command

    4263B ABORt Command :ABORt Note GPIB Reference 5-9...
  • Page 139: Calculate Subsystem

    4263B CALCulate Subsystem COMMAND PARAMETER j j j j j j j j j j j < > < > < > < > < > < > < > < > Note 5-10 GPIB Reference...
  • Page 140 4263B :CALCulate1:FORMat REAL MLINear :CALCulate2:FORMat IMAGinary PHASe REAL Note GPIB Reference 5-11...
  • Page 141 4263B Table 5-2. Measurement Parameter Choices Measurement Parameter SENSe:FUNCtion:ON :CALCulate :LIMit:BEEPer:CONDition FAlL PASS f j g Note :CALCulate :LIMit:BEEPer[:STATe] 5-12 GPIB Reference...
  • Page 142 4263B Note :CALCulate :LIMit:CLEar :CALCulate :LIMit:FAIL? :CALCulate :LIMit:LOWer[:DATA] < > numeric value numeric value < > numeric < > :CALCulate :LIMit:LOWer:STATe :CALCulate :LIMit:STATe :CALCulate :LIMit:UPPer[:DATA] < > numeric value numeric value < > numeric < > GPIB Reference 5-13...
  • Page 143 4263B :CALCulate :LIMit:UPPer:STATe :CALCulate :MATH:EXPRession:CATalog? :CALCulate :MATH:EXPRession:NAME PCNT :CALCulate :MATH:STATe :CALCulate :PATH? :CALCulate :MATH:STATe 5-14 GPIB Reference...
  • Page 144: Calibration Subsystem

    4263B CALibration Subsystem COMMAND PARAMETER < > :CALibration:CABLe < > numeric value < > Cable Length Available Frequency Range < > GPIB Reference 5-15...
  • Page 145 4263B DATA Subsystem COMMAND PARAMETER < > < > < > :DATA[:DATA] REF1 REF2 , < > numeric value 5-16 GPIB Reference...
  • Page 146 4263B :DATA[:DATA]? BUF1 BUF2 set1 set2 setn < > < > < > set1 < > set2 < > setn < > stat data comp < > < > < > stat < > data < > comp < >...
  • Page 147 4263B , < > :DATA:FEED BUF1 BUF2 data handle data handle < > :DATA:FEED:CONTrol BUF1 BUF2 ALWays NEVer :DATA:POINts BUF1 BUF2 , < > numeric value numeric value < > numeric < > 5-18 GPIB Reference...
  • Page 148 4263B DISPlay Subsystem COMMAND PARAMETER < > f j j g f j g f j j j j j g :DISPlay[:WINDow][:STATe] :DISPlay[:WINDow]:TEXT1:DIGit :DISPlay[:WINDow]:TEXT1:PAGE GPIB Reference 5-19...
  • Page 149 4263B :DISPlay[:WINDow]:TEXT2:PAGE 5-20 GPIB Reference...
  • Page 150: Fetch? Query

    4263B FETCh? Query :FETCh? stat data1 data2 comp1 comp2 < > < > < > < > < > stat < > data1 < > data2 < > comp1 < > comp2 < > GPIB Reference 5-21...
  • Page 151: Format Subsystem

    4263B FORMat Subsystem COMMAND PARAMETER :FORMat[:DATA] ASCii REAL[,64] 5-22 GPIB Reference...
  • Page 152: Initiate Subsystem

    4263B INITiate Subsystem COMMAND PARAMETER < > :INITiate[:IMMediate] :lNITiate:CONTinuous GPIB Reference 5-23...
  • Page 153: Sense Subsystem

    4263B SENSe Subsystem COMMAND PARAMETER < > < > < > < > < > < > < > < > < > < > < > [:SENSe]:AVERage:COUNt < > numeric value numeric value < > numeric < > [:SENSe]:AVERage[:STATe]...
  • Page 154 4263B [:SENSe]:CORRection:CKIT:STANdard3 < > , < > numeric value numeric value numeric value < > numeric value < > Note numeric value < > numeric < > [:SENSe]:CORRection:COLLect[:ACQuire] STANdard Setting STANdard1 STANdard2 STANdard3 (OPEN (SHORT (LOAD correction) correction) correction) Test Signal Frequency...
  • Page 155 4263B [:SENSe]:CORRection:COLLect:METHod REFL2 REFL3 [:SENSe]:CORRection:DATA? STANdard < > [:SENSe]:CORRection[:STATe] [:SENSe]:FIMPedance:APERture < > [MS numeric value numeric value < > numeric < > [:SENSe]:FIMPedance:CONTact:VERify [:SENSe]:FIMPedance:RANGe:AUTO 5-26 GPIB Reference...
  • Page 156 4263B [:SENSe]:FIMPedance:RANGe[:UPPer] < > [MOHM KOHM MAOHM] numeric value numeric value < > numeric < > [:SENSe]:FUNCtion:CONCurrent sensor function sensor function < > < > sensor function < > sensor function < > Command [:SENSe]:FUNCtion:CONCurrent [:SENSe]:FUNCtion:COUNt? sensor function < >...
  • Page 157 4263B Table 5-3. Measurement Function Selection < > < > < > < > < > < > < > < > 5-28 GPIB Reference...
  • Page 158 4263B SOURce Subsystem COMMAND PARAMETER < > < > < > < > :SOURce:FREQuency[:CW] < > [HZ KHZ] numeric value numeric value < > numeric < > :SOURce:VOLTage[:LEVel][:IMMediate][:AMPLitude] < > [MV numeric value numeric value < > numeric < >...
  • Page 159 4263B :SOURce:VOLTage[:LEVel][:IMMediate]:OFFSet:STATe 5-30 GPIB Reference...
  • Page 160 4263B STATus Subsystem COMMAND PARAMETER < > < > :STATus:OPERation[:EVENt]? < > :STATus:OPERation:CONDition? < > :STATus:OPERation:ENABle < > numeric value numeric value < > numeric < > :STATus:PRESet :STATus:QUEStionable[:EVENt]? :STATus:QUEStionable:CONDition? GPIB Reference 5-31...
  • Page 161 4263B :STATus:QUEStionable:ENABle < numeric value > numeric value < > numeric value < > 5-32 GPIB Reference...
  • Page 162 4263B SYSTem Subsystem COMMAND PARAMETER < > < > < > :SYSTem:BEEPer[:IMMediate] :SYSTem:BEEPer:STATe :SYSTem:ERRor? :SYSTem:KLOCk :SYSTem:LFRequency < > numeric value numeric value < > numeric < > :SYSTem:PRESet GPIB Reference 5-33...
  • Page 163 4263B Item Reset key Stored in :SYSTem:VERSion? 5-34 GPIB Reference...
  • Page 164 4263B TRIGger subsystem COMMAND PARAMETER < > :TRIGger:DELay < > [MS numeric value numeric value < > numeric < > < > :TRIGger[:IMMediate] :TRIGger:SOURce EXTernal INTernal MANual GPIB Reference 5-35...
  • Page 165: Common Commands

    4263B Common Commands ESE < > numeric value < > ESE? < > ESR? < > IDN? eld 1 eld 2 eld 3 eld 4 < > < > < > < > eld 1 eld 2 eld 3 eld 4...
  • Page 166 4263B numeric value < > numeric value SAV < > numeric value numeric value < > numeric value SRE < > numeric value numeric value < > numeric value SRE? < > STB? < > TST? Test Item Error Code <...
  • Page 167 4263B 5-38 GPIB Reference...
  • Page 168: Status Reporting Structure

    4263B Status Reporting Structure Figure 5-2. Status Reporting Structure Service Request (SRQ) GPIB Reference 5-39...
  • Page 169: Status Byte Register

    4263B Status Byte Register Figure 5-3. Status Byte Register Table 5-4. Status Byte Assignments Bit No. Description Weight 5-40 GPIB Reference...
  • Page 170 4263B Statndard Event Status Register Figure 5-4. Standard Event Status Register GPIB Reference 5-41...
  • Page 171 4263B Table 5-5. Standard Event Status Register Assignments Bit No. Description Weight Standard Operation Status Group 5-42 GPIB Reference...
  • Page 172 4263B Figure 5-5. Standard Operation Status Group Structure GPIB Reference 5-43...
  • Page 173: Operation Status Register

    4263B Operation Status Register Table 5-6. Operation Status Condition Register Assignments Bit No. Description Weight Table 5-7. Operation Status Event Register Assignments Bit No. Description Weight Questionable Status Register 5-44 GPIB Reference...
  • Page 174 4263B Table 5-8. Questionable Status Register Assignments Bit No. Bit Weight Description GPIB Reference 5-45...
  • Page 175: Trigger System

    4263B Trigger System 4263B Trigger System Con guration Figure 5-6. Trigger System Con guration trigger state Idle Initiate Trigger Event Detection Sequence Operation IDLE State 5-46 GPIB Reference...
  • Page 176: Initiate State

    4263B Initiate State Trigger Event Detection State Figure 5-7. Inside an Trigger Event Detection State Sequence Operation State GPIB Reference 5-47...
  • Page 177: Data Transfer Format

    4263B Data Transfer Format ASCii < > < > < > Figure 5-8. NR1 Format Figure 5-9. NR2 Format Figure 5-10. NR3 Format < > < > < > 5-48 GPIB Reference...
  • Page 178 4263B REAL Figure 5-11. REAL Data Format < > < > < > < > < > < > < < (EXP-1023) -l022 GPIB Reference 5-49...
  • Page 180: Application Measurement

    Application Measurement Measuring Electrolytic Capacitors (Sample Program) Figure 6-1. Measuring Electrolytic Capacitors Application Measurement 6-1...
  • Page 181 4263B 6-2 Application Measurement...
  • Page 182 4263B Figure 6-2. Sample Program to Measure Electrolytic Capacitors Caution Application Measurement 6-3...
  • Page 183 4263B 6-4 Application Measurement...
  • Page 184 4263B Note Application Measurement 6-5...
  • Page 185 4263B Measuring Transformers Note 6-6 Application Measurement...
  • Page 186 4263B Figure 6-3. Connecting a Transformer Application Measurement 6-7...
  • Page 187 4263B Note 6-8 Application Measurement...
  • Page 188 4263B Measuring High Inductance Transformers Application Measurement 6-9...
  • Page 190: Impedance Measurement Basics

    Impedance Measurement Basics Impedance Measurement Basics 7-1...
  • Page 191: Characteristics Example

    4263B Characteristics Example Figure 7-1. Typical Characteristics of Components 7-2 Impedance Measurement Basics...
  • Page 192: Parallel/Series Circuit Mode

    4263B Parallel/Series Circuit Mode Table 7-1. Parallel/Series Circuit Model Circuit Model Measurement De nition of D, Q, and G Function Impedance Measurement Basics 7-3...
  • Page 193 4263B Selecting Circuit Model for Capacitance Measurement Small Capacitance Values Large Capacitance Values Figure 7-2. Capacitance Circuit Model Selection 7-4 Impedance Measurement Basics...
  • Page 194 4263B Selecting Circuit Model for Inductance Measurement Small Inductance Values Large Inductance Values Figure 7-3. Inductance Circuit Model Selection Impedance Measurement Basics 7-5...
  • Page 195 4263B Four-Terminal Pair Con guration Figure 7-4. Four-Terminal Pair Measurement Principle 7-6 Impedance Measurement Basics...
  • Page 196: Measurement Contacts

    4263B Measurement Contacts Figure 7-5. Measurement Contacts Impedance Measurement Basics 7-7...
  • Page 197: Capacitance To Ground

    4263B Capacitance to Ground Figure 7-6. Model of Capacitance to Ground Figure 7-7. Reducing Capacitance to Ground 7-8 Impedance Measurement Basics...
  • Page 198: Contact Resistance

    4263B Contact Resistance Figure 7-8. Contact Con guration Extending Test Leads Figure 7-9. Measurement Contacts for Test Lead Extension Impedance Measurement Basics 7-9...
  • Page 199 4263B Using a Guard Plate for Low Capacitance Measurements Figure 7-10. Example of DUT Guard Plate Connection Shielding Figure 7-11. Guard Shield 7-10 Impedance Measurement Basics...
  • Page 200: Contact Check

    4263B Contact Check Figure 7-12. Stray Capacitance Causing Contact Check Error Impedance Measurement Basics 7-11...
  • Page 201 4263B Correction Functions of the 4263B Table 7-2. Correction Functions Correction Description Typical Usage Selection 7-12 Impedance Measurement Basics...
  • Page 202 4263B Impedance Measurement Basics 7-13...
  • Page 203 4263B Standard for the LOAD Correction Using a Standard Supplied by a Component Manufacturer Using Your Own Standard Selecting the LOAD standard. Measuring the LOAD reference value. 7-14 Impedance Measurement Basics...
  • Page 204: Speci Cations

    Speci cations Speci cations 8-1...
  • Page 205: Measurement Conditions

    4263B Speci cations Measurement Parameters Measurement Conditions < 8-2 Speci cations...
  • Page 206: Measurement Range

    4263B Measurement Range Parameter Measurement Range Parameter Measurement Range Measurement Accuracy < < < < Speci cations 8-3...
  • Page 207 4263B j j j j > 2 2 j > > Figure 8-1. Conversion Diagram 8-4 Speci cations...
  • Page 208 4263B Table 8-1. Measurement Accuracy Parameter: A, B, and C A [%] (Short / Medium, Long) B [%] (Short / Medium, Long) Test Signal Frequency 100/120 Hz 1 kHz 10 kHz j < j < j < < j < j <...
  • Page 209 4263B Table 8-1. Measurement Accuracy Parameter: A, B, and C (continued) A [%] (Short / Medium, Long) B [%] (Short / Medium, Long) Test Signal Frequency 20 kHz 100 kHz j < j < j < < j j <...
  • Page 210 4263B > < fC x fL x RS e fL x fC x Speci cations 8-7...
  • Page 211 4263B RP x RP e 8-8 Speci cations...
  • Page 212: Measurement Support Functions

    4263B Measurement Support Functions Speci cations 8-9...
  • Page 213 4263B General 8-10 Speci cations...
  • Page 214: Supplemental Performance Characteristics

    4263B Supplemental Performance Characteristics < Test Signal Level < < < < < < < < < Speci cations 8-11...
  • Page 215 4263B Parameter Measurement Range AE L AE L =L x L x =E Cable Test Signal Frequency Length 100/120 Hz 1 kHz 10 kHz 20 kHz 100 kHz Test Signal Frequency 100/120 Hz 1 kHz 10 kHz 20 kHz 100 kHz...
  • Page 216 4263B AE R AE R D =R x Cable Length D" G=L x I=L x Measurement Test Signal Frequency Time Mode 100/120 Hz 1 kHz 10 kHz 20 kHz 100 kHz Cable Length Test Signal Frequency 100/120 Hz 1 kHz 10 kHz 20 kHz 100 kHz...
  • Page 217 4263B Measurement Test Signal Frequency Time Mode 100/120 Hz 1 kHz 10 kHz 20 kHz 100 kHz Test Signal Level < < < < 8-14 Speci cations...
  • Page 218 4263B Table 8-4. Measurement Time (DC Bias: OFF) Measurement Analog Digital Tzm1 Time Mode Measurement Computation (= Analog Meas. + Dig. Comput.) Table 8-5. Measurement Time (DC Bias: ON) Measurement Test Signal Test Signal Level Tzm2 Time Mode Frequency Speci cations 8-15...
  • Page 219 4263B < Table 8-6. Additional Measurement Time for Rdc Measurement Measurement Tdcr Time Mode Table 8-7. Additional Measurement Time for N and M Measurement Measurement Time Mode approx. 10 N approx. 10 Measurement Maximum DC Range Bias Current 8-16 Speci cations...
  • Page 220 4263B < < < Figure 8-2. Maximum Capacitor Voltage Caution Speci cations 8-17...
  • Page 222: Maintenance

    Maintenance Maintenance 9-1...
  • Page 223: Test Equipment

    4263B Test Equipment Table 9-1. Recommended Test Equipment Equipment Critical Speci cations Recommended Qty. Use Model < < < 9-2 Maintenance...
  • Page 224: Performance Tests

    4263B Performance Tests Introduction Note Note Test Equipment Calculation Sheet Maintenance 9-3...
  • Page 225: Calibration Cycle

    4263B Performance Test Record Calibration Cycle How to Set the 4263B for the Performance Tests Meas Time 9-4 Maintenance...
  • Page 226: Speci Cation

    4263B Test Signal Frequency Accuracy Test Speci cation Note Test Equipment Description Recommended Model Procedure Figure 9-1. Test Signal Frequency Accuracy Test Setup Maintenance 9-5...
  • Page 227 4263B Table 9-2. Frequency Accuracy Test Settings Test Signal Frequency 9-6 Maintenance...
  • Page 228 4263B Test Signal Level Accuracy Test Speci cation Test Equipment Description Recommended Model Procedure Figure 9-2. Test Signal Level Accuracy Test Setup Shift 5 495 495 495 475 495 495 495 Enter Maintenance 9-7...
  • Page 229 4263B Table 9-3. Level Accuracy Test Settings Test Signal Test Signal Level Frequency 9-8 Maintenance...
  • Page 230 4263B DC Bias Level Accuracy Test Speci cation Test Equipment Description Recommended Model Procedure Description Recommended Model Figure 9-3. DC Bias Level Accuracy Test Setup Maintenance 9-9...
  • Page 231 4263B Figure 9-4. DC Bias Level Accuracy Test Setup Without The Interface Box Table 9-4. Bias Level Accuracy Test Settings DC Bias Level 9-10 Maintenance...
  • Page 232 4263B 0 m Impedance Measurement Accuracy Test Speci cation Test Equipment Description Recommended Model Procedure 0 m Capacitance Measurement Accuracy Test (Meas. Time Mode: LONG). Figure 9-5. 0 m Impedance Measurement Accuracy Test Setup Maintenance 9-11...
  • Page 233 4263B Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: 9-12 Maintenance...
  • Page 234 4263B Table 9-5. 0 m Capacitance Measurement Test (LONG) Settings Standard Test Signal Frequency 0 m Capacitance Measurement Accuracy Test (Meas. Time Mode: MED). Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Maintenance 9-13...
  • Page 235 4263B Table 9-6. 0 m Capacitance Measurement Test (MED) Settings Standard Test Signal Frequency 0 m Capacitance Measurement Accuracy Test (Meas. Time Mode: SHORT). Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Table 9-7. 0 m Capacitance Measurement Test (SHORT) Settings...
  • Page 236 4263B Table 9-8. 0 m Capacitance Measurement Test (DC Bias) Setting Standard Test Signal Frequency 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: LONG). Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Maintenance 9-15...
  • Page 237 4263B Table 9-9. 0 m Resistance Measurement Test (LONG) Settings Standard Test Signal Frequency 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: MED). Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Table 9-10. 0 m Resistance Measurement Test (MED) Settings...
  • Page 238 4263B Table 9-11. 0 m Resistance Measurement Test (SHORT) Settings Standard Test Signal Frequency 0 m DC Resistance Measurement Accuracy Test (Opt. 001 Only). Measurement Parameter: Measurement Time Mode: Test Signal Level: Trigger Mode: Maintenance 9-17...
  • Page 239 4263B 1 m Impedance Measurement Accuracy Test Speci cation Test Equipment Description Recommended Model Procedure 1 m Capacitance Measurement Accuracy Test. 9-18 Maintenance...
  • Page 240 4263B Figure 9-6. 1 m Impedance Measurement Accuracy Test Setup Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Maintenance 9-19...
  • Page 241 4263B Table 9-12. 1 m Capacitance Measurement Test Settings Standard Test Signal Frequency 1 m Resistance Measurement Accuracy Test. Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Table 9-13. 1 m Resistance Measurement Test Settings...
  • Page 242 4263B 1 m DC Resistance Measurement Accuracy Test (Opt. 001 Only). Measurement Parameter: Measurement Time Mode: Test Signal Level: Trigger Mode: Maintenance 9-21...
  • Page 243 4263B 2 m Impedance Measurement Accuracy Test Speci cation Test Equipment Description Recommended Model Procedure 2 m Capacitance Measurement Accuracy Test. 9-22 Maintenance...
  • Page 244 4263B Figure 9-7. 2 m Impedance Measurement Accuracy Test Setup Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Maintenance 9-23...
  • Page 245 4263B Table 9-14. 2 m Capacitance Measurement Test Settings Standard Test Signal Frequency 2 m Resistance Measurement Accuracy Test. Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Table 9-15. 2 m Resistance Measurement Test Settings...
  • Page 246 4263B 2 m DC Resistance Measurement Accuracy Test (Opt. 001 Only). Measurement Parameter: Measurement Time Mode: Test Signal Level: Trigger Mode: Maintenance 9-25...
  • Page 247 4263B 4 m Impedance Measurement Accuracy Test Speci cation Test Equipment Description Recommended Model Procedure 4 m Capacitance Measurement Accuracy Test. 9-26 Maintenance...
  • Page 248 4263B Figure 9-8. 4 m Impedance Measurement Accuracy Test Setup Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Maintenance 9-27...
  • Page 249 4263B Table 9-16. 4 m Capacitance Measurement Test Settings Standard Test Signal Frequency 4 m Resistance Measurement Accuracy Test. Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Table 9-17. 4 m Resistance Measurement Test Settings...
  • Page 250 4263B 4 m DC Resistance Measurement Accuracy Test (Opt. 001 Only). Measurement Parameter: Measurement Time Mode: Test Signal Level: Trigger Mode: Maintenance 9-29...
  • Page 251 4263B Calculation Sheet Test Signal Frequency Accuracy Test Test Signal Counter Reading Test Result Frequency Equation Test Signal Level Accuracy Test Test Signal Test Signal Multimeter Reading Test Result Level Frequency Equation DC Bias Level Accuracy Test DC Bias Multimeter Reading...
  • Page 252 4263B Standards' Calibration Values Standard Frequency Parameter Calibration Value Reference Designation Maintenance 9-31...
  • Page 253 4263B 0 m Impedance Measurement Accuracy Test 0 m Capacitance Measurement Accuracy Test (Meas. Time Mode: LONG) Standard Test Signal Parameter 4263B Reading Test Result Frequency Measured Equation 9-32 Maintenance...
  • Page 254 4263B 0 m Capacitance Measurement Accuracy Test (Meas. Time Mode: LONG) Standard Test Signal Parameter 4263B Reading Test Result Frequency Measured Equation 0 m Capacitance Measurement Accuracy Test (Meas. Time Mode: MED) Standard Test Signal Parameter 4263B Reading Test Result...
  • Page 255 4263B 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: MED) Standard Test Signal Parameter 4263B Reading Test Result Frequency Measured Equation 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: SHORT) Standard Test Signal Parameter 4263B Reading Test Result...
  • Page 256 4263B 1 m Impedance Measurement Accuracy Test 1 m Capacitance Measurement Accuracy Test Standard Test Signal Parameter 4263B Reading Test Result Frequency Measured Equation 1 m Resistance Measurement Accuracy Test Standard Test Signal Parameter 4263B Reading Test Result Frequency Measured...
  • Page 257 4263B 2 m Impedance Measurement Accuracy Test 2 m Capacitance Measurement Accuracy Test Standard Test Signal Parameter 4263B Reading Test Result Frequency Measured Equation 2 m Resistance Measurement Accuracy Test Standard Test Signal Parameter 4263B Reading Test Result Frequency Measured...
  • Page 258 4263B 4 m Impedance Measurement Accuracy Test 4 m Capacitance Measurement Accuracy Test Standard Test Signal Parameter 4263B Reading Test Result Frequency Measured Equation 4 m Resistance Measurement Accuracy Test Standard Test Signal Parameter 4263B Reading Test Result Frequency Measured...
  • Page 259: Performance Test Record

    4263B Performance Test Record Test Signal Frequency Accuracy Test Test Signal Test Limits Test Result Measurement Frequency Uncertainty Test Signal Level Accuracy Test Test Signal Test Signal Test Limits Test Result Measurement Level Frequency Uncertainty 9-38 Maintenance...
  • Page 260 4263B DC Bias Level Accuracy Test DC Bias Test Limits Test Result Measurement Level Uncertainty Record-1 Maintenance 9-39...
  • Page 261 4263B 0 m Impedance Measurement Accuracy Test 0 m Capacitance Measurement Accuracy Test (Meas. Time Mode: LONG) 1 of 2 Standard Test Signal Parameter Test Limits Test Result Measurement Frequency Measured Uncertainty 9-40 Maintenance Record-2...
  • Page 262 4263B 0 m Capacitance Measurement Accuracy Test (Meas. Time Mode: LONG) 2 of 2 Standard Test Signal Parameter Test Limits Test Result Measurement Frequency Measured Uncertainty 0 m Capacitance Measurement Accuracy Test (Meas. Time Mode: MED) Standard Test Signal Parameter...
  • Page 263 4263B 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: LONG) Standard Test Signal Parameter Test Limits Test Result Measurement Frequency Measured Uncertainty 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: MED) Standard Test Signal Parameter Test Limits Test Result...
  • Page 264 4263B 1 m Impedance Measurement Accuracy Test 1 m Capacitance Measurement Accuracy Test Standard Test Signal Parameter Test Limits Test Result Measurement Frequency Measured Uncertainty 1 m Resistance Measurement Accuracy Test Standard Test Signal Parameter Test Limits Test Result Measurement...
  • Page 265 4263B 2 m Impedance Measurement Accuracy Test 2 m Capacitance Measurement Accuracy Test Standard Test Signal Parameter Test Limits Test Result Measurement Frequency Measured Uncertainty 2 m Resistance Measurement Accuracy Test Standard Test Signal Parameter Test Limits Test Result Measurement...
  • Page 266 4263B 4 m Impedance Measurement Accuracy Test 4 m Capacitance Measurement Accuracy Test Standard Test Signal Parameter Test Limits Test Result Measurement Frequency Measured Uncertainty 4 m Resistance Measurement Accuracy Test Standard Test Signal Parameter Test Limits Test Result Measurement...
  • Page 267: Functional Tests

    4263B Functional Tests Introduction Test Equipment 9-46 Maintenance...
  • Page 268 4263B Transformer Measurement Functional Test (Opt. 001 Only) Test Equipment Description Recommended Model Procedure Figure 9-9. Transformer Measurement Functional Test Setup Measurement Parameter: Measurement Time Mode: Test Signal Level: Test Signal Frequency: Trigger Mode: Maintenance 9-47...
  • Page 269 4263B Table 9-18. Transformer Measurement Test Limits Switch Frequency Minimum Maximum Setting Setting Limit Limit 9-48 Maintenance...
  • Page 270 4263B Handler Interface Functional Test Test Equipment Description Recommended Model Procedure Initial Setup. Figure 9-10. Handler Interface Functional Test Setup Key Lock Function Test. External Trigger Function Test. Maintenance 9-49...
  • Page 271 4263B Handler Interface Output Test. Caution Figure 9-11. Handler interface Output Order 9-50 Maintenance...
  • Page 272 4263B Contact Check Functional Test Test Equipment Description Recommended Model Procedure Figure 9-12. Contact Check Functional Test Setup Maintenance 9-51...
  • Page 273 4263B 9-52 Maintenance...
  • Page 274 4263B Functional Test Record Transformer Measurement Functional Test (Opt. 001 Only) Pass Fail Maintenance 9-53...
  • Page 276: Manual Changes

    Manual Changes Introduction Manual Changes MANUAL CHANGES Table A-1. Manual Changes by Serial Number Serial Pre x Make Manual Changes or Number Manual Changes A-1...
  • Page 277: Serial Number

    4263B Serial Number Figure A-1. Serial Number Plate A-2 Manual Changes...
  • Page 278: Electrical Characteristics

    Handler Interface Installation Introduction Electrical Characteristics Output Signals Table B-1. Handler Output Electrical Characteristics Output Signals Voltage Output Rating Maximum External Voltage / Current Circuit Common High Comparison Signals Control Signals Handler Interface Installation B-1...
  • Page 279 4263B Figure B-1. Handler Interface Comparison Output Signals Diagram B-2 Handler Interface Installation...
  • Page 280 4263B Figure B-2. Handler Interface Control Output Signals Diagram Handler Interface Installation B-3...
  • Page 281: Input Signals

    4263B Input Signals Table B-2. Handler Interface Input Electrical Characteristics Signal Input Voltage Input Current (Low) Circuit Common High Pull-up Voltage 12 V 15 V Figure B-3. Handler Interface Input Signal Diagram B-4 Handler Interface Installation...
  • Page 282 Tools and Fasteners Procedure Warning Dangerous energy and voltage levels exist within the 4263B when it is in operation and just after it is powered down. Allow 10 minutes for the 4263B's internal capacitors to discharge before starting to work on it.
  • Page 283 4263B B-6 Handler Interface Installation...
  • Page 284 4263B Figure B-4. A1 Main Board Handler Interface Installation B-7...
  • Page 285 4263B Table B-3. Pull-up Resistor Location Socket No. Signal Name Signal Type Pull-up Pull-up Resistor Voltage Agilent Part Number Pull-up Pull-up Resistor Voltage Agilent Part Number B-8 Handler Interface Installation...
  • Page 286 4263B EXT DCV2 EXT TRIG KEY LOCK SW1-1 SW1-2 SW1-3 Handler Interface Installation B-9...
  • Page 288 Overload/No-Contact Operations Table C-1. OVLD/N.C. Condition Display Handler GPIB Output Solutions Output Data Mode Comprtr Mode < > < > < > < > < > < > < > < > < > Overload/No-Contact Operations C-1...
  • Page 289 4263B Table C-2. Simultaneous OVLD and N.C. Condition Condition Display Handler GPIB Output Solutions Output Data Mode Comprtr Mode < > < > < > < > < > < > < > < > < > C-2 Overload/No-Contact Operations...
  • Page 290 Messages Messages-1...
  • Page 291: Instrument Errors

    4263B Instrument Errors Messages-2...
  • Page 292: Gpib Errors

    4263B GPIB Errors -100 -101 -102 -103 -104 -105 -108 -109 -112 Messages-3...
  • Page 293 4263B -113 -121 -123 -124 -128 -131 -138 -140 -141 -144 -148 -150 Messages-4...
  • Page 294 4263B -151 -158 -160 -161 -168 -170 -171 -178 -200 -211 -213 Messages-5...
  • Page 295 4263B -221 -222 -223 -230 -241 -310 -311 -313 -350 -400 -410 -420 Messages-6...
  • Page 296 4263B -430 -440 Messages-7...
  • Page 298 Index Special characters Index-1...
  • Page 299 data handle < > Index-2...
  • Page 300 Index-3...
  • Page 301 < > data handle < > numeric value < > sensor function < > numeric value Index-4...
  • Page 302 sensor function < > Index-5...
  • Page 303 Index-6...
  • Page 304 REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative.
  • Page 305 Agilent 4263B LCR Meter Operation Manual Manual Change Agilent Part No. N/A Jun 2009 Change 1 Add TAR in Test Signal Frequency Accuracy Test (Page 9-38) as follows. Test Signal Frequency Accuracy Frequency Limits [Hz] Result [Hz] Uncertainty [Hz] [Hz] ±0.01...
  • Page 306 1 μF ±0.0011 ± 0.000032 34.29 1 μF 10 k ±0.0026 μF μF ± 0.00014 μF 19.45 1 μF 10 k ±0.0026 ± 0.000069 38.02 1 μF 100 k ±0.0176 μF μF ± 0.00018 μF 102.72 Copyright 2009 Agilent Technologies ○...
  • Page 307 Limits Result Uncertainty 17.76 100 mΩ ±0.52 mΩ mΩ ± 0.030 mΩ 16.31 100 mΩ ±0.48 mΩ mΩ ± 0.030 mΩ Note 1: Result = Measurement – Calibrated Value of Standard Measurement: Resistance Cable length: Copyright 2009 Agilent Technologies ○...
  • Page 308 Status PASS Standard Parameter Limits Result Uncertainty ± 0.039 25.06 100 mΩ ±0.97 mΩ mΩ mΩ 67.56 100 kΩ ±0.87 kΩ mΩ ± 0.013 kΩ Note 1: Result = Measurement – Calibrated Value of Standard Copyright 2009 Agilent Technologies ○...
  • Page 309 PASS Frequency [Hz] Standard Limits Result Uncertainty ± 0.030 35.88 100 mΩ ±1.05 mΩ mΩ mΩ 26.19 100 kΩ ±0.87 kΩ mΩ ± 0.014 kΩ Note 1: Result = Measurement – Calibrated Value of Standard Copyright 2009 Agilent Technologies ○...
  • Page 310 Test signal level: Status PASS Standard Parameter Limits Result Uncertainty 38.28 100 mΩ ±1.13 mΩ mΩ ± 0.030mΩ 66.20 100 kΩ ±0.87 kΩ mΩ ± 0.014 kΩ Note 1: Result = Measurement – Calibrated Value of Standard Copyright 2009 Agilent Technologies ○...
  • Page 311 Status PASS Standard Parameter Limits Result Uncertainty ± 0.030 43.74 100 mΩ ±1.29 mΩ mΩ mΩ 68.26 100 kΩ ±0.87 kΩ kΩ ± 0.013 kΩ Note 1: Result = Measurement – Calibrated Value of Standard Copyright 2009 Agilent Technologies ○...
  • Page 312 Agilent 4263B LCR Meter Operation Manual Manual Change Agilent Part No. N/A July 2008 Change 1 Change the Measurement Uncertainty value of Test Signal Frequency Accuracy Test (Page 9-38) to the following information. Test Signal Test Limits Test Result Measurement...
  • Page 313 ± 0.128 nF ± 0.00087 nF ±0.0128 ± 0.00022 0.1 μF 1 kHz ± 0.11 nF ± 0.0068 nF ±0.0011 ± 0.000023 0.1 μF 100 kHz ± 1.47 nF ± 0.015 nF ±0.0147 ± 0.00026 Copyright 2007 Agilent Technologies ○...
  • Page 314 Measurement Standard Frequency Measured Test Limits Test Result Uncertainty 0.01 μF 100 Hz ± 0.179 nF ± 0.011 nF ±0.0179 ± 0.00072 0.01 μF 100 kHz ± 0.287 nF ± 0.00081 nF ±0.0287 ± 0.00069 Copyright 2007 Agilent Technologies ○...
  • Page 315 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: SHORT) (Page 9-42) to the following information. Test Signal Parameter Measurement Standard Frequency Measured Test Limits Test Result Uncertainty 100 mohm 100 Hz ±1.10 mohm mohm ± 0.13 mohm Copyright 2007 Agilent Technologies ○...
  • Page 316 1 m Resistance Measurement Accuracy Test (Page 9-43) to the following information. Test Signal Parameter Measurement Standard Frequency Measured Test Limits Test Result Uncertainty 100 mohm 100 Hz ±0.60 mohm mohm ± 0.030 mohm 100 mohm 1 kHz ±0.60 mohm mohm ± 0.031 mohm Copyright 2007 Agilent Technologies ○...
  • Page 317 2 m Resistance Measurement Accuracy Test (Page 9-44) to the following information. Test Signal Parameter Measurement Standard Frequency Measured Test Limits Test Result Uncertainty 100 mohm 100 Hz ±0.68 mohm mohm ± 0.030 mohm 100 mohm 1 kHz ±0.72 mohm mohm ± 0.034 mohm Copyright 2007 Agilent Technologies ○...
  • Page 318 4 m Resistance Measurement Accuracy Test (Page 9-45) to the following information. Test Signal Parameter Measurement Standard Frequency Measured Test Limits Test Result Uncertainty 100 mohm 100 Hz ±0.84 mohm mohm ± 0.030 mohm 100 mohm 1 kHz ±0.96 mohm mohm ± 0.030 mohm Copyright 2007 Agilent Technologies ○...
  • Page 319 4 m DC Resistance Measurement Accuracy Test (Opt. 001.Only) (Page 9-45) to the following information. Measurement Standard Test Limits Test Result Uncertainty 100 mohm ±1.29 mohm mohm ± 0.030 mohm 100 kohm ±0.87 kohm kohm ± 0.013 kohm Copyright 2007 Agilent Technologies ○...
  • Page 320 Agilent 4263B LCR Meter Operation Manual Manual Change Agilent Part No. N/A July 2007 Change 1 Add the following steps to the step 2 of DC Bias Level Accuracy Test (page 9-9). 2-a. Set the test signal frequency to 100 kHz using Freq.
  • Page 321 マニュアル チェンジ 変更 1 DC バイアス・レベル確度試験の手順 2(ページ 9-9)に以下の手順を追加して下さい。 2-a. Freq を押して測定信号周波数を 100 kHz に設定します。 2-b. Level を押して測定レベルを 50 mV に設定します。 変更 2 DC バイアス・レベル確度試験用セットアップの図 9-4(ページ 9-10)を以下に変更して下さい。 Copyright 2007 Agilent Technologies ○...
  • Page 322 June 2007 Change 1 Add the following note to the “3-c. Turning On the 4263B and Setting the Line Frequency” (page 1-13). Note In case the setting of the ac line frequency is not match, there’s any possibility the measurement error will rise.
  • Page 323 マニュアル チェンジ 変更 1 電源投入と電源周波数の設定の手順 3-c(ページ 1-11)に以下の注記を追加して下さい。 注記 AC 電源周波数の設定値が一致していない場合、測定誤差が増大する可能性があります。必ず AC 電源周波数 を正しく設定して下さい。 変更 2 測定確度の条件(ページ 8-3)に以下の条件を追加して下さい。 5. AC 電源周波数が正しく設定されている。 Copyright 2007 Agilent Technologies ○...
  • Page 326: Safety Summary

    When you notice any of the unusual conditions listed below, immediately terminate operation and disconnect the power cable. Contact your local Agilent Technologies sales representative or authorized service company for repair of the instrument. If you continue to operate without repairing the instrument, there is a potential fire or shock hazard for the operator.
  • Page 330 Do not apply DC voltage or current to the UNKNOWN terminals. Do not apply External DC bias voltage more than +2.5V. Doing so will damage the 4263B. Before you measure a capacitor, be sure the capacitor is fully discharged. 4263B...

Table of Contents