11

Regular maintenance tasks

Check the cathode
buffer fill level
156
Regular maintenance tasks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156
Manage the cartridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158
Install cathode buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 166
When a cartridge is installed in the instrument, the volume of buffer in the Cathode
Buffer Container (CBC) must be above the fill line.
Check the cathode buffer fill level before each run. It is recommended to check one
time each week if the instrument is not in use.
In the home screen:
1. Touch
, touch
2. Press the release button on the autosampler to open the lid, then remove the
CBC.
Release button
1
CBC
2
3. Ensure that the level of buffer is
above the fill line.
If the buffer is at or below the fill
line, see "Assemble the SeqStudio
Genetic Analyzer Cathode Buffer
Container (CBC)" on page 166 and
"Insert the Cathode Buffer
Container" on page 167.
Maintain the instrument
Eject plate, then open the instrument door when prompted.
2
1
1
SeqStudio
Genetic Analyzer Instrument and Software User Guide
Replace if buffer is at or below the fill
line
1