Investigating RF/DC Issues

RF dip is incorrect.

Possible Remedy
Make sure the RF board is properly connected and use the Power Status tab on the ISQ 7000 Dashboard to ensure the readback reports
+48 V DC is present. If +48 V DC is not present, contact Technical Support.

RF/DC System Check diagnostic fails.

Possible Remedies
Use ISQ 7000 Tune to tune the RF system. If the RF system will not tune, contact Technical Support.
Make sure the rod driver board is properly installed and undamaged. Replace it, if necessary. See
details.
Thermo Scientific
When you run RF/DC System Check diagnostics on the ISQ 7000 system, the software
checks the RF/DC system of the instrument. Your instrument or software may indicate the
following types of issues.
WARNING ELECTRICAL SHOCK HAZARD: When troubleshooting any issue that requires
removing a cover on the ISQ 7000 mass spectrometer, you should power-off and vent the
instrument to avoid harming yourself.
AVERTISSEMENT RISQUE D'ÉLECTROCUTION : lorsque vous cherchez à résoudre un
problème nécessitant d'enlever le couvercle du spectromètre de masse ISQ 7000, mettez
hors tension et ventilez l'instrument pour éviter de vous blesser.
2
Troubleshooting
Investigating RF/DC Issues
RF Board and Rod Driver Board
ISQ 7000 Hardware Manual