Capillary
Cleaning and/or
Replacement
450iQ Instruction Manual
Use the following procedure to clean or replace the capillary.
Equipment required:
Phillips drive, #2
Hex drive, 7/64
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2–1).
2. Disconnect the plumbing.
3. Unfasten captive hardware.
Figure 7–25. Remove Capillary, Disconnect and Unfasten
4. Slide capillary plate upwards clearing the partition panel keyway.
Thermo Scientific