Servicing

Capillary Cleaning and/or Replacement

Capillary
Cleaning and/or
Replacement
7-22
48iQ Instruction Manual
Use the following procedure to clean and/or replace the capillary.
Equipment required:
Phillips drive, #2
Hex drive, 7/64
1. Turn instrument OFF, unplug power cord, and remove the cover
(Figure
2–1).
2. Disconnect the plumbing.
3. Using #2 Phillips drive, unfasten captive hardware.
Figure 7–25. Remove Capillary, Disconnect and Unfasten
4. Slide capillary plate upwards clearing the partition panel keyway.
Thermo Scientific