2.10 Connection of Coolant

2.10.1 Install water chiller

Follow the instructions supplied with
the chiller.
Connect up the water chiller to the
push-fit ports on right hand side of
instrument (taking care to connect
the output connection from the
chiller to the input connection on the
instrument, after removing the
shipment plugs).
Make sure the supplied water filter
is fitted onto the input to the
instrument and does not leak (check
tightness of filter cover).
Fill the water chiller with coolant
The coolant should be distilled, or
deionised water, with 2.5% v/v corrosion
inhibitor (842312052621).
(This volume equates to 190ml for a
ThermoFlex900 chiller with standard 3m
pipes).
The inhibitor is included with the instrument
and as part of the Installation and
Maintenance Kit.
(Full coolant information is included in the
Pre-Installation guide).
WARNING: do not use any
other cooling liquid mixture.
Note: Do Not switch on the chiller until starting the performance tests. This is
to avoid ice damage to the CID camera if water has not been removed from the
optical system.
Note: When disconnecting push fit connectors - first push in the sleeve as far as possible as
this will then release the tubing.
Reminder: Soak the spray chamber in a 10% laboratory cleaning solution (for
example Decon) for 30 minutes. This will save time later in the installation.
Remember to remove the o-rings.
Page 8 of 14
Installation Details