ERROR/FAULT MESSAGES
The instrument displays error/fault messages relating to analysis by indicators on the
display or by flag screens.
Display Indicators
A letter or symbol at the end of the analyzer reading indicates the following:
Symbol
H
High alarm level exceeded
L
Low alarm level exceeded
S
Short Term Exposure Limit alarm level exceeded
X
Fault/error
P
Purge not done
Z
Zero invalid
>
Concentration exceeds full scale value (HRL).
Measurements are below the range of the measurement electronics. Possible
corrective action is to rezero the analyzer or change the pathlength.
Measurements exceed the range of the measurement electronics. Possible
corrective action is to rezero the analyzer or change the pathlength.
Calculation overflow where range has been exceeded for processing a
}
measurement value. Possible corrective action is to repeat the measurement.
Table 4-2. Display Indicators
Description