Advanced Troubleshooting
Replacing Components of the Ion Source

Repeller Plate

198
TSQ 9610 Hardware Manual
To replace the repeller plate:
Note See the ISQ and TSQ GC-MS Spare Parts Guide for information about ordering this
component.
1. Shut down your system by following the steps in
on
page
139.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note Remove the screws completely so that they do not scratch the manifold when you
close the door.
4. Open the manifold door.
Figure 194. Opening the Manifold Door
"Shutting Down the TSQ 9610 System"
Screws