Spectrum Analyzer Commands
[:SENSe]:BANDwidth|BWIDth[:RESolution]:RATio <number>
[:SENSe]:BANDwidth|BWIDth[:RESolution]:RATio?
Title: Resolution Bandwidth to Span Ratio
Description: Sets the ratio of the resolution bandwidth to the span for use when the
Parameter: <number>
Default Value: 0.01
Range: 0.00001 to 1
Related Command: :BANDwidth|BWIDth[:RESolution]:AUTO
Front Panel Access: BW, Span/RBW (note that this is the inverse ratio)
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration <freq>
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration?
Title: Channel Power Integration Bandwidth
Description: Sets the integration bandwidth for the channel power measurement.
Parameter: <freq>
Default Value: 10.35 MHz
Default Unit: Hz
Related Command: :FREQuency:SPAN
Front Panel Access: Shift-4 (Measure), Channel Power, Ch Pwr Width
[:SENSe]:CHPower:STATe OFF|ON|0|1
[:SENSe]:CHPower:STATe?
Title: Channel Power Measurement State
Description: Sets the state of the channel power measurement, ON or OFF. When
Parameter: OFF|ON|0|1
Parameter Type: <boolean>
Default Value: OFF
Related Command: :CONFigure:ACPower
Front Panel Access: Shift-4 (Measure), Channel Power, On/Off
S412E PM
resolution bandwidth to span coupling is enabled. Note that the front
panel interface sets the inverse ratio: the span to the resolution
bandwidth.
Integration bandwidth must be less than or equal to the frequency span.
using :CONFigure:CHPower, the state is automatically set to ON.
PN: 10580-00319 Rev. N
4-14 [:SENSe] Subsystem
4-57