Fixed WiMAX Commands
[:SENSe]:FREQuency:SIGStandard:NAMe <string>
[:SENSe]:FREQuency:SIGStandard:NAMe?
Title: Signal Standard
Description: Selects the desired signal standard from the list. The <string>
Parameter: <string>
Front Panel Access: Freq, Signal Standard
[:SENSe]:PFail <test set>
[:SENSe]:PFail?
Title: PASS/FAIL Test Selection
Description: Selects the active test set to be used in subsequent PASS/FAIL
Parameter: <test set>
Related Command: :CONFigure PFail
Front Panel Access: Measurements, Pass/Fail Mode, Select Pass/Fail Test
S412E PM
argument is the name of the desired signal standard as displayed in the
instrument's current signal standard list. The list can be displayed on
the instrument by choosing the Signal Standard submenu button in the
Freq menu. The list can also be downloaded remotely and viewed using
Anritsu Master Software Tools. For example, if the desired Signal
Standard is P-GSM 900 - Uplink, the value of the <string> argument
would be "P-GSM 900 - Uplink".
The query form of this command will return the name of the currently
selected Signal Standard on the list.
measurements. The <test set> value must correspond to a test set that
is defined in the test set list. To view the list on the instrument, go into
the "Measurement" menu, press the "Pass/Fail Mode" key, and then
press the "Select Pass/Fail Test" submenu. Alternatively, the list can be
retrieved remotely and viewed using the Anritsu Master Software
Tools. The <test set> argument is a 1-based index of the position of the
desired Pass/Fail test set in the instrument's current Pass/Fail test set
list. For example, if the desired Pass/Fail test is the 3rd item on the list,
the value of the <test set> argument would be 3.
When using the query form of the command, the return value is the
currently selected test set number. If there is not a valid test set
selected, the return value is "-1" (negative 1).
PN: 10580-00319 Rev. N