NBFM Commands
:MEASure:SIGAnalyzer?
Description: Sets the active measurement to NBFM Analyzer, triggers a new
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To measure NBFM Analyzer numerical data:
Front Panel Access: NA
S412E PM
measurement and returns the numerical results. It is a combination of
the commands :CONFigure:SIGAnalyzer and :READ:SIGAnalyzer?
Data is returned as 10 comma-separated values in the following order
and format:
Carrier Power (dBm as float or Watts as long long int or Volts as
long long int)
Carrier Frequency (Hz as float)
Deviation (Hz as float)
Mod Rate (Hz as float)
SINAD (dB as float)
THD (% as float)
Occ BW (Hz as float)
ExtSINAD (dB as float)
CTCSS (string)
DCS (string)
DTMF (string)
If there is no valid measurement data, the instrument will return
"--,--,--,--,--,--,--,--,--,--,--".
Please note that the received power will be returned in the unit that is
selected through the Rx Units button on the front panel or with the
command :UNIT:POWer:RX. If the receiver unit has been set to dBm,
the received power is returned in dBm. If the unit is set to Watts, the
received power is returned in fW. If the unit is set to Volts, the received
power is returned in fV.
:MEASure:SIGAnalyzer?
PN: 10580-00319 Rev. N
16-7 :MEASure Subsystem