Spectrum Analyzer Commands
:MEASure:CHPower?
Title: Measure Channel Power/Density
Description: Sets the active measurement to channel power, sets the default
Default Unit: Current amplitude units
Related Command: :MEASure:CHPower:CHPower?
:MEASure:OBWidth?
Title: Measure Occupied Bandwidth
Description: Sets the active measurement to occupied bandwidth, sets the default
Default Unit: OBW in Hz, Percent of Power in %, dB Down in dB
Related Command: :CONFigure:OBWidth
S412E PM
measurement parameters, triggers a new measurement and returns the
channel power and channel power density results. It is a combination of
the commands :CONFigure:CHPower and :READ:CHPower? For a
description of the default channel power measurement parameters, see
:CONFigure:CHPower. To make a channel power measurement with
settings other than the default values, send:
:CONFigure:CHPower
Commands to set desired settings
:READ:CHPower?
Data is returned as 2 comma-separated values: channel power, channel
power density.
:MEASure:CHPower:DENSity?
:CONFigure:CHPower
measurement parameters, triggers a new measurement and returns the
occupied bandwidth, percent of power and dB down results. It is a
combination of the commands :CONFigure:OBWidth and
:READ:OBWidth? For a description of the default occupied bandwidth
measurement parameters, see :CONFigure:OBWidth. To make an
occupied bandwidth measurement with settings other than the default
values, send:
:CONFigure:OBWidth
Commands to set desired settings
:READ:OBWidth?
Data is returned as 3 comma-separated values: occupied bandwidth,
percent of power, dB down.
:CONFigure:RF SPECtrum