Spectrum Analyzer Commands
:CALCulate:LIMit:TYPe 0|1
:CALCulate:LIMit:TYPe?
Title: Set Limit Line Upper or Lower
Description: Sets the currently active limit line to either upper or lower. Subsequent
Parameter: 0|1 (0 = upper limit line, 1 = lower limit line)
Parameter Type: number
Default Value: 0 (upper)
Range: 0|1
Example: To set the active limit line to upper:
Related Command: :None
Front Panel Access: Shift-6 (Limit), Limit
:CALCulate:LIMit:UPPer:ALARm OFF|ON|0|1
:CALCulate:LIMit:UPPer:ALARm?
Title: Upper Limit Alarm
Description: Enables/disables the alarm for the upper limit. It is a combination of the
Parameter: OFF|ON|0|1
Parameter Type: <boolean>
Default Value: OFF
Related Command: :CALCulate:LIMit:ALARm
Front Panel Access: Shift-6 (Limit), Limit Alarm
:CALCulate:LIMit:UPPer:FAIL?
Title: Upper Limit Fail State
Description: Query whether the upper limit line has failed or not. The command
S412E PM
limit line operations will be performed on the selected limit line.
:CALCulate:LIMit:TYPe 0
commands :CALCulate:LIMit:TYPe 0 and :CALCulate:LIMit:ALARm
ON|OFF. Setting the value to ON or 1 will turn on the upper limit
alarm. Setting the value to OFF or 0 will turn off the upper limit alarm.
The query version of the command returns a 1 if the upper limit line
alarm is ON and returns 0 if OFF.
returns a 0 on success, 1 on fail, and 2 if the upper limit line is OFF or
the alarm is OFF.
PN: 10580-00319 Rev. N
4-2 :CALCulate Subsystem
4-9