[:Sense]:Frequency Subsystem - Anritsu LMR Master S412E Programming Manual

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

5-11 [:SENSe] Subsystem

[:SENSe]:FREQuency Subsystem

Commands in this subsystem pertain to the frequency settings of the instrument.
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
Description: Sets the center frequency. Note that changing the value of the center
Cmd Parameter: <NRf> <freq> (hertz)
Query Response: <NR3> <freq> (hertz)
Default Value: For 1.6 GHz Model: 800250000 Hz
Default Unit: Hz
Range: For 1.6 GHz Model: 500 kHz to 1.6 GHz
Front Panel Access: Freq/Dist (or Freq), Center Freq
[:SENSe]:FREQuency:SPAN <freq>
[:SENSe]:FREQuency:SPAN?
Description: Sets the frequency span. Setting the value of <freq> to 0 Hz is the
Cmd Parameter: <NRf> <freq> (hertz)
Query Response: <NR3> <freq> (hertz)
Default Value: For 1.6 GHz Model: 1599500000 Hz
Default Units: Hz
Range: For 1.6 GHz Model: 0 Hz to 1.599500 GHz
Front Panel Access: Freq/Dist (or Freq), Span
5-76
frequency will change the value of the coupled
parameters: Start Frequency and Stop Frequency. It may also
change the value of the span.
For 6 GHz Model: 3000250000 Hz
For 6 GHz Model: 500 kHz to 6 GHz
equivalent of setting the span mode to zero span. Note that changing
the value of the frequency span will change the value of coupled
parameters: Start Frequency and Stop Frequency, and may change the
Center Frequency.
For 6 GHz Model: 5999500000 Hz
For 6 GHz Model: 0 Hz to 5.999500 GHz
PN: 10580-00319 Rev. N
VNA Commands
S412E PM

Advertisement

Table of Contents
loading

Table of Contents