16-9 :READ Subsystem
:READ:SIGAnalyzer?
Description: Triggers a new NBFM Analyzer measurement and returns the
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To read NBFM Analyzer numerical data:
Related Command: :CONFigure:SIGAnalyzer
Front Panel Access: NA
16-26
numerical results. It is a combination of the commands :ABORT;
:INITiate; :FETCh:SIGAnalyzer?
NBFM Analyzer must be the active measurement (specified by
:CONFigure:SIGAnalyzer). The current measurement can be queried
using :CONFigure?
Data is returned as 10 comma-separated values in the following order
and format:
Carrier Power (dBm as float or Watts as long long int or Volts as
long long int)
Carrier Frequency (Hz as float)
Deviation (Hz as float)
Mod Rate (Hz as float)
SINAD (dB as float)
THD (% as float)
Occ BW (Hz as float)
ExtSINAD (dB as float)
CTCSS (string)
DCS (string)
DTMF (string)
If there is no valid measurement data, the instrument will return
"--,--,--,--,--,--,--,--,--,--,--".
Please note that the received power will be returned in the unit that is
selected through the Rx Units button on the front panel or with the
command :UNIT:POWer:RX. If the receiver unit has been set to dBm,
the received power is returned in dBm. If the unit is set to Watts, the
received power is returned in fW. If the unit is set to volts, the received
power is returned in fV.
Note: This command is not affected by the squelch level set using the
front panel.
:READ:SIGAnalyzer?
PN: 10580-00319 Rev. N
NBFM Commands