VNA Commands
[:SENSe]:CORRection:COLLect:ACQUire:STATus? [<cal steps>,
<port_no>]
Description: This command requests information about the current calibration step
Cmd Parameter: NA (query only)
Query Parameter: <char> [<cal steps>, <port_no>]
Query Response: <NR1> <integer>
Front Panel Access: NA
[:SENSe]:CORRection:COLLect:CONNector<port_no> <connector>
[:SENSe]:CORRection:COLLect:CONNector<port_no>?
Description: Sets the connector family for the given port number. <port_no> is
Cmd Parameter: <char> <connector>
Query Response: <char> <connector> (returns short format only)
Default Value: KMAL
Front Panel Access: Shift-2 (Calibrate), DUT Port Setup, DUT Port 1
[:SENSe]:CORRection:COLLect:MEDium COAX
[:SENSe]:CORRection:COLLect:MEDium?
Description: Sets the calibration line type.
Cmd Parameter: <char> COAX
Query Response: <char> COAX
Default Value: COAX
Front Panel Access: Shift-2 (Calibrate)
S412E PM
or the specified calibration step. If no calibration step is specified, then
it returns a 1 if the current calibration step has completed, otherwise it
returns a 0. <cal steps> is the calibration step to be performed and
must be one of the following values:
OPEN|SHORT|LOAD|THRU|ISOLation
<port_no> is port number 1. For step OPEN, SHORT, valid port
number is 1. For calibration step THRU and ISOLation, valid port
number is 1 for Fwd.
port number 1. <connector> defines the connector family. Valid
connector for calibration line type COAX is the following:
NMALe|NFEMale|KMALe|KFEMale|716Male|716Female|TNCMale|
TNCFemale|SMAMale|SMAFemale|USR1|USR2|USR3|USR4
Note that the connector must be valid for the current calibration line
type.