VNA Commands
5-7

:MMEMory Subsystem

The Mass Memory subsystem contains functions that provide access to the instrument setup
and data storage.
:MMEMory:LOAD:STATe <integer>, <file name>
Description: Recalls a previously stored setup from the current save location. The
Cmd Parameter: <integer>, <string> (1, file name)
Query Response: NA (no query)
Related Command: :MMEMory:STORe:STATe
Front Panel Access: Shift-7 (File), Recall
S412E PM
saved setup that is to be loaded is specified by <file name>.
<file name> must be enclosed in either single quotes (' ') or double
quotes (" ") and must include the extension ".stp". The <integer>
parameter is not currently used, but it must be sent. Send a value of 1.
:MMEMory:MSIS INTernal|USB