[:SENSe]:POWer[:RF]:RANGe[:IMMediate]
Title: Amplitude Range
Description: Re-calculates amplitude range. Note that issuing this command will set
Related Command: [:SENSe]:POWer[:RF]:RANGe:AUTO
Front Panel Access: Amplitude, Adjust Range
[:SENSe]:POWer[:RF]:RANGe:AUTO <Boolean (default=1 [ON])>
[:SENSe]:POWer[:RF]:RANGe:AUTO?
Title: Automatic Amplitude Range
Description: Sets the automatic amplitude range. Setting the value to ON or 1 will
Parameter: OFF|ON|0|1
Parameter Type: <boolean>
Default Value: ON
Related Command: [:SENSe]:POWer[:RF]:RANGe
Front Panel Access: Amplitude, Auto Range
[:SENSe]:RF:SPECtrum:SPAN Auto|1.4|3|5|10|15|20|30
[:SENSe]:RF:SPECtrum:SPAN?
Description: Sets the span for the Spectrum view. For example, setting the value to 5
Parameter: Auto|1.4|3|5|10|15|20|30
Default Value: Auto
Default Unit: MHz
Front Panel Access: Measurements, RF, Channel Spectrum, Span
S412E PM
the automatic dynamic range OFF.
result in the amplitude range being coupled to the detected input signal
level. Setting the value to OFF or 0 will result in the input attenuation
being uncoupled from the input signal level. That is, changing the input
signal level will not change the amplitude range. When this command is
issued, the amplitude range itself will not change. The default value is
ON. That is, sending :SENS:POW:RANG:AUTO is equivalent to
sending :SENS:POW:RANG:AUTO ON.
will set the span for the Spectrum view to 5 MHz and setting the value
to 20 will set the span for the Spectrum view to 20 MHz.