Anritsu LMR Master S412E Programming Manual page 295

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

P25 Phase 1 Commands
:FETCh:SIGAnalyzer?
Description: Returns the most recent P25 Analyzer numerical measurement results.
Cmd Parameters: NA
Query Parameters: NA
Range: NA
Default Value: NA
Default Unit: NA
Example: To fetch P25 Analyzer numerical data:
Related Command: :CONFigure:SIGAnalyzer
Front Panel Access: NA
S412E PM
Data is returned as 7 comma-separated values in the following order
and format:
Received Power (dBm as float or Watts as long long int or Volts as
long long int)
Frequency Error (Hz as float)
Mod Fid (% as float)
BER (% as float)
Symbol Dev (Hz as float)
NAC (hex)
Symbol Rate Error (Hz as float)
If there is no valid measurement data, the instrument will return
"--,--,--,--,--,--,--".
The received power will be returned in the unit that is selected through
the Rx Units button on the front panel or with the command:
UNIT:POWer:RX. If the receiver unit has been set to dBm, the received
power is returned in dBm. If the unit is set to Watts, the received power
is returned in fW (10
is returned in fV (10
:FETCh:SIGAnalyzer?
PN: 10580-00319 Rev. N
-15
W). If the unit is set to Volts, the received power
-15
V).
10-4 :FETCh Subsystem
10-9

Advertisement

Table of Contents
loading

Table of Contents