5-2 :CALCulate Subsystem

:CALCulate:MARKer Subsystem

This subsystem contains commands to manipulate data markers.
:CALCulate:MARKer:AOFF
Turns off all markers. This command will turn off all the markers and set Marker Type to off.
:CALCulate:MARKer:DATA?
Description: Reports the marker information. Each marker data is separated by a
Cmd Parameter: NA (query only)
Query Response: <char> (comma separated data)
Front Panel Access: NA
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa:REFerence <Mk>
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:DELTa:REFerence?
Description: Sets the specified delta marker reference to the given reference marker
Cmd Parameter: <char> <Mk>
Query Response: <char> <Mk>
Default Value: –230 (The selected marker is a reference marker)
Example: Set Marker 1 as the reference marker of delta Marker 3:
Related Command: :CALCulate:MARKer<Mk>:TYPE,
Front Panel Access: Marker, Avail Ref Mkr
5-22
comma and data are returned similar to that when Readout Format is
set to Table.
specified by <Mk>. <Mk> is the reference marker number in the range
of 1 to 8. The query version of the command returns the reference
marker number to which the specified delta marker should be
referenced. If the selected marker is not a delta marker, then –230 is
returned. Note that the set version of this command will set the
specified delta marker as the active marker. The given reference
marker number must be currently set as a reference marker and the
specified delta marker number must currently be set as delta marker.
Also both markers (delta and reference) must be in the same domain
type.
:CALCulate:MARKer3:DELTa:REFerence 1
:CALC:MARK3:DELT:REF 1
:CALCulate:MARKer<Mk>:DOMain?
PN: 10580-00319 Rev. N