Anritsu LMR Master S412E Programming Manual page 284

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

9-11 [:SENSe] Subsystem
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer?
Title: Center Frequency
Description: Sets the center frequency.
Parameter: <freq>
Default Value: 1 GHz
Default Unit: Hz
Range: 10 MHz to 4 GHz
Front Panel Access: Freq, Center Freq
[:SENSe]:FREQuency:SIGStandard:CHANnel <number>
[:SENSe]:FREQuency:SIGStandard:CHANnel?
Title: Channel Selection
Description: Sets the channel number for the selected signal standard.
Parameter: <number>
Front Panel Access: Frequency, Channel
[:SENSe]:FREQuency:SIGStandard:NAMe [String]
[:SENSe]:FREQuency:SIGStandard:NAMe?
Title: Signal Standard
Description: Selects the desired signal standard from the list. The <string>
Parameter: <string>
Front Panel Access: Freq, Signal Standard
[:SENSe]:LTE:STATus?
Description: This command queries the LTE sweep complete status. Returns 1 when
Cmd Parameter: NA
Query Response: 0 | 1
Related Command: :STATus:OPERation?
Front Panel Access: NA
9-24
argument is the name of the desired signal standard as displayed in the
instrument's current signal standard list. The list can be displayed on
the instrument by choosing the Signal Standard submenu button in the
Freq menu. The list can also be downloaded remotely and viewed using
Anritsu Master Software Tools. For example, if the desired Signal
Standard is P-GSM 900 - Uplink, then the value of the <string>
argument would be "P-GSM 900 - Uplink".
The query form of this command will return the name of the currently
selected Signal Standard from the list.
the sweep is complete. Returns 0 when the sweep is in progress.