System Subsystem - Anritsu LMR Master S412E Programming Manual

An integrated, handheld multi-function land mobile radio test tool
Hide thumbs Also See for LMR Master S412E:
Table of Contents

Advertisement

3-6 :SYSTem Subsystem

3-6
:SYSTem Subsystem
This subsystem contains commands that affect instrument functionality. This functionality
does not directly relate to data collection, display, or transfer.
:SYSTem:MSIS[:STATe]? INTernal|USB
Title: Query Memory State
Description: Queries the ready state of the instrument's internal memory or of the
Parameter: INTernal|USB
Parameter Type: <char>
Related Command: :MMEMory:MSIS
:SYSTem:OPTions?
Title: Query Installed Options
Description: Returns a string of the installed options. Options are separated by a "/".
Cmd Parameter: NA (query only)
Query Response: NA (options are separated by "/" or "NONE")
Related Command: *IDN?
:SYSTem:PRESet
Title: Preset
Description: This command restores all application parameters to their factory
Syntax: :SYSTem:PRESet
Cmd Parameter: NA
Query Response: NA (no query)
Related Command: *RST
Front Panel Access: Shift-1 (Preset), Preset
3-10
USB Flash drive. Use this command to check the ready state of the
memory device before sending a command, such as :MMEMory:STORe
or :MMEMory:MSIS:COPY, that requires the memory location to be
available.
The USB query returns a 1 when a USB device is plugged into the USB
port. It returns 0 if no USB drive is present.
Internal memory in the LMR Master S412E should always be available
and the INT query should always return a 1.
:MMEMory:MSIS:DESTination
:MMEMory:MSIS:COPY
The string returns "NONE" if no options are installed.
preset values. This command does not modify system parameters such
as Ethernet configuration, language, volume, or brightness.
Note: After issuing this command, wait 30 seconds for the instrument
to complete the preset condition before sending another command.