NBFM Commands

16-12 :UNIT Subsystem

The unit subsystem is used to modify the default units used for related parameters. These
changes affect parameters in both commands and responses.
:UNIT:POWer:RX DBM|WATT|VOLTs
:UNIT:POWer:RX?
Description: Sets the receiver unit to dBm or Watts or Volts. If the unit is set to dBm,
Cmd Parameters: DBM|WATT|VOLTs
Query Parameters: NA
Range: DBM|WATT|VOLTs
Default Value: WATT
Default Unit: NA
Example: To set the receiver units to watts:
Front Panel Access: Amplitude, Units, Rx Units
:UNIT:POWer:TX DBM|WATT|VOLTs
:UNIT:POWer:TX?
Description: Sets the generator unit to dBm or volts. If the unit is set to dBm, the
Cmd Parameters: DBM|WATT|VOLTs
Query Parameters: NA
Range: DBM|WATT|VOLTs
Default Value: VOLTs
Default Unit: NA
Example: To set the generator units to volts:
Front Panel Access: Amplitude, Units, Tx Units
S412E PM
the NBFM Analyzer received power (from FETCh:SIGAnalyzer? or
READ:SIGAnalyzer? or MEASURE:SIGAnalyzer?) and the squelch
setting will be set and queried in dBm.If the unit is set to Watts, the
NBFM Signal Analyzer received power and squelch setting will be set
and queried in fW. If the unit is set to Volts, the NBFM Signal Analyzer
received power and squelch setting will be set and queried in fV.
:UNIT:POWer:RX WATT
Tx Output Lvl setting will be set and queried in dBm. If the unit is set
to Watts, the Tx Output Lvl setting will be set and queried in fW. If the
unit is set to Volts, the Tx Output Lvl will be set and queried in fV.
:UNIT:POWer:TX VOLT